LG 17HQ701G-B Owner's Manual page 32

Flat panel digital x-ray detector
Hide thumbs Also See for 17HQ701G-B:
Table of Contents

Advertisement

Connect - Wired Connection
Use the LAN cable to connect a PC to the Control Box and connect the detector to the Control Box with the main cable.
After connecting the cables, proceed to set up the PC as shown below.
1
Launch the [Network and Sharing Center] and click [Change adapter settings].
- Control Panel > Network and Internet > Network and Sharing Center > Change adapter settings
2
Go to [Properties] in Local Area Connection.
3
Select [Internet Protocol Version 4 (TCP/IPv4)] and then click [Properties] to set the IP address as follows:
- IP address: 10.10.10.2 ~ 10.10.10.254 (set an IP address other than 10.10.10.99, 10.10.10.100)
- [Subnet Mask]: 255.255.255.0
- [Default Gateway]: 10.10.10.1
- Do not set up DNS.
32
4
Launch Calibration Software, go to
(Settings) > [DXD] > [Connection & File Save] and enter the detector's
IP (10.10.10.100). Click the [Ping] button to check the connection.
Press the [Ping] button. If connected successfully, the following pop-up will appear: