Trigger Cable
Trigger Cable is connected between control box and X-ray Generator, and used only for manual mode, not auto mode.
DXD 1
DXD 2
Ethernet
Sync
AC-IN
T4L 250V
No.
Labeling of pins color
1
Red
Power: X-ray Generator Supply Voltage (DC 12 V ~24 V)
2
Brown
Expose signal from Generator to Control Box
3
Blue
Cancel REQ Ground
4
Orange
Prepare Acknowledge signal from Control Box to Generator
5
Yellow
Expose Acknowledge signal from Control Box to Generator
6
Green
Prepare signal from Generator to Control Box
7
Black
Cancel request VCC
8
Gray
Prepare Acknowledge Ground
9
Violet
Ground of Signals
NOTE
Trigger Cable and X-ray Generator connection is to be performed by qualified personnel. Description of each pin is common language of this industry.
NC: No Connection
Description
Use
Use
NC
Use
Use
Use
NC
NC
Use
Labeling of pins
X-ray Generator