Druck ADTS 2XX Calibration Manual
2.3
Sensor Calibration of the Pitot and Static Channels
Before starting this procedure carry out a calibration check of all channels.
Set the unit to Leak/Measure mode, set operating limits to MAX and disable Auto Leak,
refer to the User manual for further details.
Set the calibration switch to enable (switch latched above the front panel level).
Note: Pressure values other than those stated in Table 2-3 can be used within the ranges of Ps1, Ps2 and Pt. A
minimum of two calibration points must be used.
Procedure
(1)
To perform a calibration, select SETUP and Calibrate Sensors:
(2)
Adjust the pressure standard and apply pressure No. 1 in Table 2-3, allow the displayed reading
to stabilize.
(3)
Press, Enter Pt reference and, using the numeric keys, set the value of pressure No. 1. The
screen shows the number of points entered Pt = 1.
K0425 Issue No. 2
6