Table of Contents

Advertisement

LOGIQ 7/LOGIQ 7 Pro Quick Guide

Using Probes

Connecting a probe
1.
Place the probe's carrying case on a stable surface
and open the case.
2.
Carefully remove the probe and unwrap the probe
cable.
3.
DO NOT allow the probe head to hang free. Impact
to the probe head could result in irreparable damage.
4.
Turn the connector locking handle clockwise.
5.
Align the connector with the probe port and carefully
push into place.
6.
Turn the connector locking handle clockwise to
secure the probe connector.
7.
Carefully position the probe cable in the probe cord
holder spot so it is free to move, but not resting on
the floor.
Activating the probe
Select the appropriate probe from the probe indicators on
the Touch Panel.
The probe activates in the currently-selected operating
mode. The probe's default settings for the mode and
selected exam are used automatically.
Using Probes
Direction 5307393-100 Rev. 1
Deactivating the probe
When deactivating the probe, the probe is automatically
placed in standby mode.
Press the Freeze key or select another probe.
Disconnecting the probe
Probes can be disconnected at any time. However, the
probe should not be selected as the active probe.
1.
Move the probe locking handle counterclockwise.
Pull the probe and connector straight out of the
probe port.
2.
Carefully slide the probe and connector away from
the probe port and around the right side of the
keyboard. Ensure the cable is free.
3.
Be sure that the probe head is clean before placing
the probe in its storage box.
14

Advertisement

Table of Contents
loading

This manual is also suitable for:

Logiq 7

Table of Contents