Warnings ............................2 Cautions .............................3 Note ..............................3 Contraindications ..........................3 Symbols Glossary ..........................4 How to Contact Philips Respironics ...................5 System Overview ..........................5 Available System One Heated Humidifiers ................6 System Features and Contents ....................6 Connecting the Therapy Device ....................7 Connecting the Optional Heated Tubing ................. 8 Disconnecting the Optional Heated Tubing ................
Warnings A warning indicates the possibility for injury to the user or the operator. • Use the humidifier only for its intended use as described in this manual. Use only with Philips Respironics full-face masks, nasal masks, and connectors. •...
Philips Respironics, may affect the performance of the product. Note This humidifier is an optional accessory to Philips Respironics therapy device and can only operate when it is connected with the therapy. Refer to the therapy device user manual for important EMC information and additional warnings and cautions.
Symbols Glossary The following symbols may appear on this device and its packaging. Refer to http://www.symbols. philips.com for a description of the symbols used on this device and its packaging. ymbol efinition...
Should you experience trouble with this equipment or require assistance, please contact your home care provider. If you need to contact Philips Respironics directly, contact customer service at 1-800-345-6443 (USA or Canada only) or go to www.respironics.com to find your local customer...
Flexible Heated Tubing (optional) The optional heated tube connects from the humidifier to the patient’s mask. Humidifier Connector End Connect this end of the tubing to the humidifier. Note: This humidifier is specified for use with Philips Respironics REMstar and BiPAP therapy devices only.
Connecting the Therapy Device 1. When using a humidifier, attach the patient circuit to the air outlet port. 2. To connect the therapy device to the humidifier, remove the side cover on the therapy device. Press in the tab on the side cover and pull the cover away from the unit. Note: Therapy device shown is an example.
Air Outlet Port Tabs Note: Non-heated Philips Respironics tubing (15 or 22 mm) can still be used with this humidifier with the heated tubing air outlet port. Simply connect the tubing to the air outlet port. Refer to the therapy device manual for setup instructions.
Disconnecting the Therapy Device CAUTION: To avoid spilling, do not disconnect the humidifier from the therapy device with water in the tank. Remove the water tank from the humidifier before disconnecting the therapy device. 1. Grasp the release latch on the bottom of the humidifier. 2.
Page 12
3. Press the tab in the hole on top of the tank in toward the front of the tank. Gently remove the tank lid from the tank base and set aside. Empty any remaining water from the base of the tank. Hinges Tank Lid Middle Seal...
Page 13
5. Reassemble the tank by placing the hinges on the tank lid over the 2 tabs on the back of the tank base. Close the lid until the tab on the lid snaps back under the lip in the tank base. Inspect the tank.
Cleaning and Disinfection Instructions Summary ome cleaning isinfection ospital leaning ospital isinfection Humidifier Wipe with a lint-free Wipe with a lint-free Wipe with a lint-free Wipe with DisCide® Ultra Towelettes*. Enclosure cloth dampened with cloth dampened cloth dampened with (Exterior mild liquid dishwash- with 70% Isopropyl mild liquid dishwash-...
ome cleaning isinfection ospital leaning ospital isinfection Humidifier Immerse and agitate Immerse and agitate Immerse and agitate Immerse in a 75±2°C bath of sterile Door Seal in a solution of mild in 70% Isopropyl in a commercially deionized water for 30 minutes. liquid dishwashing Alcohol.
Page 16
Outlet Port Interior 1. Clean the humidifier outlet port by using a lint-free cloth dampened with a solution of mild dish washing detergent (e.g. Dawn Ultra) and potable water. A ratio of 1 teaspoon of Dawn per gallon (3.8 liters) of water is required. Insert the cloth approximately 7 cm (2.75 inches) into the outlet opening while cleaning.
Page 17
Humidifier Door Seal, Dry Box, Dry Box Seal and Inlet Seal 1. Open the humidifier lid with the release lever, and then slide the water tank out of the humidifier enclosure. 2. Remove the dry box assembly by following the instructions under “Water Tank and Lid, Humid- ifier Door Seal, Dry Box Seal, Dry Box, Inlet Seal and Heated Tubing”.
5. Allow the device to air dry completely before plugging in the power cord. 6. Inspect the device, and all circuit parts for damage after disinfection. If any parts are dam- aged, contact Philips Respironics Customer Service. Replace any damaged parts. Outlet Port Interior Complete the following steps to disinfect weekly using 70% Isopropyl Alcohol.
7. Leave the humidifier lid open and allow to air dry. 8. Reinstall the water tank and the humidifier door seal. 9. Inspect the air outlet port for damage or wear (cracking, crazing, etc.) . Discard and replace the humidifier if inspection of the air outlet port shows damage or wear. Humidifier Door Seal, Dry Box, Inlet Seal, Dry Box Seal, Water Tank and Tank Lid The humidifier door seal, dry box, inlet seal, dry box seal, water tank and tank lid may be disinfected every 3 days using 70% Isopropyl Alcohol.
Page 20
Outlet Port Interior CAUTION: Follow all instructions from the manufacturer of the enzymatic detergent. Any deviation from these or the manufacturer’s instructions, including using agents not listed in this guide, may impact the performance of the product. Review all applicable instructions for additional warnings and cautions.
Institution Cleaning section for instructions. CAUTION: Only the hospital and institution cleaning and disinfection procedures listed in this manual are recommended by Philips Respironics. Use of other cleaning and disinfection procedures, not specified by Philips Respironics, cannot be guaranteed to maintain the performance or durability of the product.
Page 22
6. Allow the device to air dry completely before plugging in the power cord. 7. Inspect the device, and all circuit parts for damage after disinfection. If any parts are damaged, contact Philips Respironics Customer Service. Replace any damaged parts.
Dispose of this device in accordance with local collections and recycling regulations. For more information, visit www.philips.com/recycling. Service The humidifier does not require routine servicing. If any part of the humidifier is worn or damaged, contact Philips Respironics or your home care provider. See the “Troubleshooting” section in this manual for additional information.
If high leak persists, issue may be caused by a misaligned or damaged seal. Seals are not user-serviceable components. Please contact your home care provider or Philips Respironics. The water tank is not Remove the water tank from the humidifier enclosure, and...
Page 25
Remove the heated tubing and check for damage. If it is connected properly not damaged, reconnect the heated tubing. If the problem or is damaged. continues, contact your home care provider or Philips Respironics. The heater plate is Check the power connections to the humidifier. If the not heating.
Page 26
Route the tubing behind the bed’s headboard. If the humidifier does not operate correctly after falling, contact your home care provider or Philips Respironics. Note: For information on troubleshooting your therapy device, see the manual included with your therapy device.
IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment EN ISO 8185 General Requirements for Humidification Systems Electrical (When the heated humidifier is used with a Philips Respironics therapy device) AC Power Consumption : 100 – 240 VAC, 50/60 Hz, 2.1 A...
Humidifier Run Time Between Water Refills CPAP Mode Bi-level Mode Continuous Operation: 8 Hours 8 Hours Temperature: 20° C (68° F) 20° C (68° F) Humidity: 15% RH 15% RH Humidifier Setting: Patient Pressure: 20 cm H 25 cm H Note: Using standard ISO 17025 leak rate.
Philips Respironics may charge an evaluation fee and return shipping. What Philips Respironics will do: If a Product fails to conform to the warranties set forth above during the applicable warranty period, Philips Respironics will repair or replace the Product or refund the original purchase price, in Philips Respironics sole discretion.
Page 30
Guarantees Act 1993 (NZ). 3. Philips warrants that the products shall be free from defects of workmanship and materials, and will perform in accordance with the product specifications for a period of two (2) years from the date of purchase from an authorised Philips Homecare Provider.
Page 31
1095654 1095420 R12 JM 05/03/2022 EN-INTL...