Specifications; Storage Conditions - Philips Amara SE Instructions For Use Manual

Full face mask
Table of Contents

Advertisement

Available languages

Available languages

Specifications

Warning: The technical specifications of the mask are provided for your healthcare
professional to determine if it is compatible with your CPAP or bi-level therapy device. If
used outside these specifications, or if used with incompatible devices, the mask may be
uncomfortable, the seal of the mask may not be effective, optimum therapy may not be
achieved, and leak, or variation in the rate of leak, may affect device function.
Pressure Flow Curve
60
50
40
30
20
10
0
0
5
Resistance
Drop in Pressure at
All sizes
Deadspace Volume
P
238 ml
S
245 ml
M
256 ml
L
267 ml
Sound Levels
A-weighted Sound Power Level
A-weighted Sound Pressure Level at 1 m
Disposal
Dispose of in accordance with local regulations.

Storage Conditions

Temperature: -4º to 140º F (-20º to +60º C)
Relative Humidity: 15% to 95% non-condensing
6
Pressure Flow Curve
Amara SE Full Face Mask
10
15
Pressure cm H
O
2
50 SLPM
0 cm
H
O
2
20
25
30
100 SLPM
0.2 cm
H
2
19 dBA
11 dBA
O