Download Print this page

Philips PM 5132 Instruction Manual page 8

Function generator 0.1hz-2mhz
Hide thumbs Also See for PM 5132:

Advertisement

Available languages

Available languages

1.2.2.
Characteristic
Adjustments
Frequency indication
Setting error
Vernier frequency adjustment
Temperature coefficient
Short term drift
Long term drift
Signal output
Connection
Impedance
Load capability
Wave forms
Duty cycle
Open-circuit voltage
— setting range
maximum
value
DC (offset) voltage
— button PUSH FOR ZERO pulled,
open circuit voltage
— button PUSH FOR ZERO or
WAVE FORM button DC pressed
Attenuation
— continuous
— fixed
Distortion (sinewave)
Linearity (triangular wave)
linear
— seven range-pushbuttons
— dia! with linear scale
— knob for fine-adjustment
— linear scale on the dial
42%
+0,2 % of full scale
+5 % of the dial setting
<0,1%/K
ranges I! - VI
<0,15 %/K ranges | - VII
<0,15%
ranges Il - VI
<0,20%
ranges |- Vilwithin
15 min.
<0,25 % within 7 hours
BNC socket
50 92/600 2, selected by pushbutton
short-circuit proof
rv
sinewave
Av triangular wave
TL squarewave
MW positive squarewave
ur— negative squarewave
DC d.c. voltage without a.c.
+15 V
-10 to +10 V, continuously adjustable
<50 mV
<0,5
% in ranges | to V
<1
%inrange VI
<3
%inrange
VII
<1%

Advertisement