Performance Specification - Sanyo MDF-U730 Service Manual

Biomedical freezer
Hide thumbs Also See for MDF-U730:
Table of Contents

Advertisement

Items
Specification
Internal temp.control range | -20°C — -30°C(AT : 5°C ~ 35°C, RH80%less)
Settable Range
Temp. set range : -15°C — -35°C
Temperature sensor
Thermistor
Temperature display
LED digital display (red colored) by 1°C.
Range : -40°C — 0°C
In case Internal Temp higher/lower than set temp. 10degrees,
High/Low
ALARM lamp brinks and buzzer sounds intermittently (about
Temperature
15 min. delay)
Remote alarm contact: turns ON with buzzer sounds
Alarm
ALARM
lamp brinks, buzzer sounds intermittently, remote
Power failure
alarm contact turns ON and display is extinguished.
Remote alarm terminal 3P: contact capacity DC30v, 2A,
Remote
N.O.
-COM.
(During temperature alarm, power failure alarm, alarm test: ON)
ALARM
lamp
BUZZER:
Buzzer key
ALARM
TEST: Alarm test key
Control panel
SET: To switch the SV and PV
»» : To shift the set digit
4
: To change the SV
DEF: Defrost key
Press >> key for five seconds
"When temperature sensor is in fault,
In case sensor shorted, error code E01/HI are displayed.
In case sensor opened, error code E01/LO are displayed.
"When defrost sensor is in fault,
Self diagnosis functi
9
Unetion
In case sensor shorted, error code E02/ internal temp. are
displayed.
In case sensor opened, error code E02/ internal temp. are
displayed
Power switch
Side on rear box
Battery switch
Behind on front cover
Power supply
Local voltage
Performance specification
115V/60hz
110V/60hz
Cooling capacity
-30°C
Temperature control range
-20°C ~ 30°C
Motor rated power consumption
300W
Motor rated current
2.9 A
-3-