2285048 - Base Plate Assembly - GE Innova 2121-IQ Biplane Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare
All GE supplied vascular system floor preparation are contained in catalog. There are some
additional gantry/table mounting kits based on each mounting method: through bolts or floor
anchors.
Base plate assembly (mandatory) 2285048 – Refer to
If there is no integrated baseplate plan (2359194) delivered with S18101SG, please refer to
drawings 2359194 (Angio configuration
Illustration
Assembly and separation (Select kit) 2285050 – Refer to tables,
Above grade, Floor anchors without insert – 2285051.
Through bolts without insert – 2285646.
On grade, Floor anchors – 2286398.
Above grade, Floor anchors and insert – 2285052.
Through bolts and insert – 2285053.
Floor plate / Base plate Assembly – 2285632.
Templates (select kit) – 2285054 – Refer to
Angio/Cardio/Neuro templates – 2285056 (containing 2185979, 2127792 and 2360133 Innova
2000 and Innova 2100–IQ/3100/4100 templates).

2.3.1 2285048 - Base Plate Assembly

Table 5-7:
Item
136
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Illustration 5-33: Anchor kit choice
5-35). Please order locally the integrated base plate based on this plan..
Name
Part #
Baseplate
2285059
Lift Plate
2290939
Table 5-7
Illustration
5-34) or 2359194 (Cardio configuration
Section 2.4
Description
12 Hole Floor Baseplate
Lifting Straps
Direction 5177951-4-1EN, Revision 1
Section 2.3
Quan.
1 pc
2 pc
2 Mounting Requirements
Notes