Height:
Depth:
Weight:
Power:
Electrical:
Ground:
Video:
Fixing:

1.2.8.3 System Cable Information

GE Healthcare provides 24 m (78'9") system cables for suspension. The length of cables
dedicated on the rail cable track is recommended as 3.8 m (12'6"). After leaving cable track,
system cables go to technical room through the conduit on the ceiling, ECG monitor VGA and
power cable go to the EKG system through another conduit. From entrance of System Control
Cabinet to their correspondent connectors, the length of the cables shall be 1.3 m (4'3'').

1.2.9 Description of the DL and associated devices

An Optima System uses the DL Digital Imaging system.
The System Control Cabinet contains:
1. DL Computer,
2. Firewall
Chapter 2 Equipment Requirements
min.: 300 mm (12")
min.: 30 mm (2")
min.: 5 kg (11 lbs)
No external power supply
IEC 320 connector connected to Optima system.
Screw-on connector
VGA connector
VESA 100 mm x 100 mm (4" x 4") connection, in the middle on the backside
Illustration 2-10: Ergo suspension system cable routing scheme
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
max.: 350 mm (14")
max.: 120 mm (4")
max.: 7 kg (15.4 lbs)