Cleaning the
Clean the electrodes and patient cables with a soft cloth
Electrodes and
moistened with a recommended disinfectant or cleaning
Cables
agent from the following list:
Cetylcide@ (may discolor cable)
Cidex@
Lysol@ Disinfectant
Lysol@ Deoderizing Cleaner (may discolor cable)
Dial@ Liquid Antibacterial Soap
ammonia
409@ (may discolor cable)
10% solution of Clorox@ in water (may discolor cable)
Murphy@ Household Cleaner, or
Ves-phene 118.
Wring any excess moisture from the cloth before
cleaning.
Caution
$
I Do not clean the patient cable with alcohol. Alcohol
can cause the plastic to become brittle and may cause
the cable to fail prematurely.
n
Do not autoclave the cable or use ultrasonic cleaners.
n
The patient cable is not immersible.
m Do not use abrasive materials to clean metal
surfaces-scratches on them can cause artifacts.
n
Avoid wetting the connectors.