GE Healthcare
REV 1
SECTION 1
ENVIRONMENTAL REQUIREMENTS/LIMITATIONS
1–1
Room climate
TABLE 2-1
ENVIRONMENTAL REQUIREMENTS/LIMITATIONS
HUMIDITY
In-Use
(equipment packed)
Min.
Max.
Min.
30%
75%
10%
1–2
Equipment heat output
CHAPTER 2 – ROOM REQUIREMENTS
Storage
In-Use
Max.
Min.
95%
10°C
50°F
IN OPERATION: 1.5 KW (4950 BTU/h) during 5 sec. max.
IN STAND–BY: 500 W (1650 BTU/h).
TEMPERATURE
Storage
(equipment packed)
Max.
Min.
Max.
40°C
- 20°C
70 °C
104°F
68°F
158°F
2–1
SENOGRAPHE DMR+
pim 5128706–1–100
ALTITUDE
In-Use
Storage
+3000 m
+3000 m
9842'
9842'