Honeywell HPB User Manual page 22

Precision barometer
Hide thumbs Also See for HPB:

Advertisement

Pressure
(hPa)
1116
1114
1112
1110
Deadband
(±2.4 hPa)
1108
1106
1104
1102
1100
1.0
Pressure
(hPa)
1116
1114
1112
1110
Deadband
1108
1106
1104
1102
1100
1.0
Pressure
(hPa)
1140
1135
1130
1125
1120
1115
1110
1105
2
1
1100
1.0
Deadband - Center
HPB pressure output
Actual pressure
3.0
Deadband - Spread
HPB pressure output
Actual pressure
3.0
Idle Count
7
4
5
6
3
2.0
I = M5
IC = 0
Setting DS=40 sets the deadband to
±40 x 0.005% FS = ±0.20%FS.
Then setting DS=C1 selects the center
deadband option with a 10x multiplier.
For a 1200 hPa full scale device, this
results in a ±2.4 hPa deadband window.
5.0
Time (sec)
I= M5
IC = 0
Setting DS=40 sets the deadband to
±40 x 0.005% FS = ±0.20%FS.
Then setting DS=S1 selects the spread
deadband option with a 10x multiplier.
For a 1200 hPa full scale device, this results
in a ±2.4 hPa deadband window.
5.0
Time (sec)
12
9
8
10
11
3.0
Time (sec)
DS=40C1
RR = 0 OP=A
DS=40S1
RR = 0 OP=A
IC = 11 (skip 11)
I = M2
RR = 0
IC = 4 (skip 4)
OP = A
IC = 2 (skip 2)
IC = 0 (skip none)
Integration time
HPB pressure output
Actual pressure