Performing The 2000 Hour Overhaul - Siemens E382 EH81E 061 01 03 02 Service Manual

Table of Contents

Advertisement

Compressor Compact
E382 EH81E 061 01 03 02
Performing the 2000 hour
overhaul
1. Replace the two dust filters.
2. Using a 5 mm hexagonal wrench,
unscrew the two screws holding the front
cover and open the cover to access the
inlet filter.
3. Replace the filter located in the
compressor inlet filter housing.
4. Close and secure the compressor's front
cover.
5. Perform a functional check. See
instructions in the Operating Manual,
chapter Functional check.
Siemens-Elema AB