Philips simplyFLO User Manual page 15

Transportable oxygen concentrator
Table of Contents

Advertisement

Available languages

Available languages

simplyFLO User Manual
Visual, Audio
Indicators
A sequence of
3 long beeps
+
1 short beep
that repeats every
4 seconds�
12
Description
No Flow Alarm
This alarm occurs when
the device detects there
is no oxygen flowing in
the patient cannula� This
alarm can be silenced
by pressing the Silence
button�
Alarm Symbol
This symbol appears
when an event occurs
that generates an
audible alarm�
Alarm Silenced
Symbol
This symbol appears
when you have pressed
the alarm silence button
on the control panel to
silence an audible alarm�
Attention
This symbol appears
when corrective action
is required�
What To Do
Check the cannula
for kinks or other
obstructions that are
stopping oxygen flow
through it�
Refer to the
associated alarm
indicator also being
displayed�
Refer to the
associated alarm
indicator also being
displayed�
Ensure that the intake
and exhaust vents
are not blocked�
Allow the device to
warm up� If the alarm
continues, contact
your home care
provider�

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Respironics simplyflo

Table of Contents