Aisys CS²

Fresh gas usage

3-16
Setting fresh gas controls
Use the Fresh Gas Controls selection to set the gas control style to
O2% with Total Flow or to individual gas flow.
Selecting O2% shows O2% as the first quick key and Total Flow
l/min as the second quick key.
Selecting Flow shows balance gas l/min as one of the quick keys
and O2 l/min as the other quick key.
Note
Both O2% and Flow are available if User is selected by the Super
User. If User is not selected by the Super User, either O2% or Flow
will be unavailable.
Note
Fresh Gas Controls cannot be accessed during a case.
1.
Select System Setup - Screen Setup.
2.
Select More Settings.
3.
Select Fresh Gas Controls.
4.
Select O2% or Flow from the drop-down menu.
When set to O2%, balance gas adjusts automatically when
either the O2% or the Total Flow is changed using the gas quick
keys.
When set to Flow, balance gas and the O2% are controlled
individually using the gas quick keys.
5.
Select Close.
Use Fresh Gas Usage to view the volume of O2, Air, N2O, and
agents used for the three most recent cases.
Data only shows for gases available on the system.
Agent data shows the three most recently used agents.
1.
Select System Setup - Fresh Gas Usage.
2.
Select Case Start Time to select the patient case to view.
3.
Select Close.