Preparing For Use; Measuring Temperature - Philips DL8740 User Manual

Hide thumbs Also See for DL8740:

Advertisement

Warning
-
Do not measure the body temperature at the
ear site if the ear is inflamed or infected.
-
Stop using the device in the occurrence of any
pain. It is likely to damage the ear canal.
-
If the measured body temperature exceeds
100.4°F / 38°C and the person whose
temperature is being measured feels
uncomfortable, contact your physician.
-
It is recommended not to use the device if you
suffer from ear disease, including otitis externa
and tympanitis.
-
Keep the device out of the reach of children
and pets to avoid inhalation or swallowing of
small parts. Children may not be able to use the
device according to the instructions in this user
manual. It is not a toy.
-
When the device is used to measure the
temperature of a child, it has to be operated
by an adult. Adults can measure their own
temperature.
-
Do not throw disposable batteries into fire.
Batteries may explode.
-
Do not put the device in a wet ear canal after
swimming or bathing. This may cause injury to
the ear canal.
-
Do not use the device if it is broken or
damaged. Using a broken or damaged device
may cause injury.
-
Do not use the thermometer under temperature
extremes ( below 50.0 °F / 10.0
C or over
°
104.0°F / 40.0
°
C) or humidity extremes
(below 15%RH or over 85%RH). If you do,
this may cause inaccurate measurements.
-
Do not expose the thermometer to temperature
extremes (below -4°F / -20°C or over
140°F / 60°C) or humidity extremes (below
15%RH or over 85%RH). If you do, this may
cause inaccurate measurements.
-
Do not store the device in extreme temperature
(below -4°F / -20°C or over 140°F / 60°C) or
in extreme levels of humidity (below 15%RH
or over 85%RH). If you do, this may cause
inaccurate measurements.
Caution
-
Only use the device as described in this user
manual. Do not use it for any other purpose.
-
The device is not intended as a substitute for
consulting a doctor. Measurement results are
for reference only.
-
If you clean the ear, wait 5-10 minutes before
you use the device.
-
Earwax in the ear canal may cause inaccurate
measurements. Make sure the ear canal is clean
before you use the device.
-
Do not take apart, repair or change any part
of the device at any time, except for replacing
batteries.
-
Keep the infrared sensor dry, clean and
undamaged to ensure accurate measurements.
-
Do not touch or blow at the infrared sensor.
A dirty infrared sensor may cause inaccurate
measurements.
-
Never clean the device with an abrasive
cleaner, thinner, benzene or immerse the device
in water or other liquids.
-
When the temperature of the storage area
differs from the temperature of the measuring
area, wait at least 30 minutes for the device to
acclimatize before you use the device.
-
If you do not intend to use the device for a
long period of time, it is advised to remove
the batteries before storing. This is to prevent
possible damage due to leakage of the
batteries. If the batteries leak, remove them
carefully and replace with new batteries.
SAVE THESE
INSTRUCTIONS
Compliance with standards
-
The device meets the relevant standards for this
type of Class II medical device for home use.
-
This Philips device complies with all applicable
standards and regulations regarding exposure
to electromagnetic fields and complies with
IEC 60601-1-2
-
Skin contact parts (ABS, TPU, metal) comply
with ISO 10993-5 and ISO 10993-10.
Cytotoxicity, sensitization and irritation test
results meet the requirements.
-
Based on the current science and technology,
other potential allergic reactions are unknown.
Clinical accuracy:
Subject age
A1
A2
B`
C
group
Operation
Adjusted mode
mode
Measuring
Ear
site
Reference
Ear
body site
Clinical bias
-0.020 -0.033 -0.014 -0.016
(Δcb)
Limits of
0.203
0.195
0.198
0.199
agreement
(LA)
Clinical
0.0770
repeatability
(σr)
Display
Symbol
Description
Explanation
History icon
Display shows last
measurement result
when switching on the
device.
Bluetooth®
The device uses
Smart icon
Bluetooth® Smart for
communication.
Battery
Battery symbol
symbol
indicating status of
battery.
Battery status indications
Battery full
To indicate that the battery is fully charged the
battery symbol shows a full battery (Fig. 3).
Battery low
To indicate that the battery is low, the battery
symbol shows a nearly empty battery (Fig. 4).
When using 2x 1.5V AAA alkaline batteries,
approximately 500 measurements can be realized
(73.4°F ± 9°F, 50 ± 20% RH)

Preparing for use

The batteries (2x 1.5V AAA) are included with the
device. Slide the battery cover down to open the
battery compartment. Remove the battery strip from
the battery compartment before first use (Fig. 15).
Slide the battery cover back until it locks into place
(Fig. 5).
Pairing the thermometer to your mobile
device
Your Philips ear thermometer is equipped with
Bluetooth
®
. Download the Philips HealthSuite
health app from the App Store or Google Play.
Use the search term: Philips HealthSuite health
app. The Philips health app is available for iOS
8.0+ and Android
TM
4.4+.
1
Download the Philips HealthSuite health app
on your mobile device, start the Setup wizard
and follow the steps to create a user profile and
add the thermometer.
2 Make sure the Philips health app is active and
Bluetooth on your mobile device is on when
pairing is in progress.
3 Press the power button once, to switch on the
thermometer (Fig. 6).
-
The app identifies the thermometer and
requests to pair.
-
The Bluetooth icon lights up when the
thermometer is connected to your mobile
device.
-
When the thermometer has paired successfully
with your mobile device, your personal
measurement results are automatically
transmitted to your mobile device via Bluetooth.
4 The thermometer switches off automatically
after 30 seconds of no activity.
Note: Only when the Philips HealthSuite health
app is active, your personal health data can be
transmitted.
Cleaning the infrared sensor
To achieve accurate measurements, it is important
to check if the infrared sensor is clean. Use a cotton
swab to clean the infrared sensor.
See chapter Cleaning and storage for instructions.
After cleaning, wait approximately 5-10 minutes
before you use the device.
Cleaning the ear
Earwax in the ear canal may cause inaccurate
measurements. Make sure the ear canal is clean
before you use the device.
Carefully clean your ear canal. If you clean your ear,
wait for 5-10 minutes before you use the device.
Resetting the thermometer
If you press the °F / °C selection button on the side
of the thermometer for longer than 10 seconds,
all measurements are deleted from the device and
all settings return to the factory values.

Measuring temperature

Tips for proper measurement
External factors can influence your body
temperature. In the following cases, wait at least
30 minutes before taking your temperature:
-
When you have been lying on your ear
-
When you had your ears covered
-
When you have been exposed to very low or
very high temperatures
-
When you have been swimming or bathing
-
If you wear earplugs or hearing aids, remove
them first
®
We do not advise you to use the device in the
following circumstances as this may cause
inaccurate measurements:
-
When the temperature of the storage area
differs from the temperature of the measuring
area. In this case, wait at least 30 minutes for
the device to acclimatize before you use the
device.