...6.2.5 Current Output Page
Continued from previous page.
OP2 Cal Hold
Enter
OP2 Law
Enter
OP2 FSD
– – –
or
Enter
Lin
Log
OP2 Zero
or
Enter
Output 2 Calibration Hold
Yes
Current Output 2 can be held during calibration, if required. Select 'YES' or 'NO'.
No
or
Store.
Advance to next parameter.
Output 2 Law
Log
Current Output 2 can be either logarithmic or linear. Select 'Log' or 'Lin'.
Lin
or
Store.
Advance to next parameter.
Output 2 Full Scale
Current Output 2 full scale range: 0.5 to 2000 mg l
is 1 decade.
Set the required concentration value for Current Output 2 full scale.
Store.
Advance to next parameter.
Output 2 Zero
– –
Set the required concentration value for Current Output 2 zero.
Note. If the Output 2 Law is linear (Lin), this parameter is omitted and the zero
current output is automatically set to '0'.
Store.
Advance to next parameter.
Continued on next page.
6 PROGRAMMING...
–1
ammonia, i.e. minimum span