Operation
34
of 106
Operation
◆ Turn the main switch counterclockwise beyond position (I), until the display
switches on. Then let the switch turn back to (I).
◆ Pre-position the MOBILETT.
◆ Release the arm system and lift up the tube assembly.
◆ Place the cassette at the required position.
◆ Switch on the collimator light.
◆ Check that the light field is parallel with the sides of the cassette, adjust by
turning the MOBILETT and the collimator.
◆ Set the appropriate film focus distance, using the built-in measuring tape.
◆ Place the detachable protractor on the cassette and read the inclination of the
cassette. Adjust the tube assembly.
◆ Collimate the radiation field for the desired image size and select exposure
parameters.
◆ Release the exposure.
Observe radiation safety regulations,
(→
Page 13)