Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
• Maintenance Protocol (SPB7- 420.832.20...) • Plus MammoReport Quality Control Manual (SPB7-420.621.20 ...) Time required Approximately 2 - 2.5 hours for one CSE. Siemens Medical Solutions Register 8 SPB7-420.831.20 Page 1 of 2 MammoReport Plus Erlangen, Germany Rev. 01 01/05...
Quality Check Image Quality Check System Software Maintenance The work steps in this document that are identified by these abbreviations should be checked off in the Maintenance Protocol (SPB7-420.832.20...). MammoReport Plus Register 8 SPB7-420.831.20 Page 2 of 2 Siemens Medical Solutions Rev.
Listen to the noise of the fans. Does it sound ok? • Shut down the operating system properly and turn off the computer and any external devices. • Disconnect the power supply cable. Siemens Medical Solutions Register 8 SPB7-420.831.20 Page 1 of 2 MammoReport Plus Erlangen, Germany Rev.
Clean the housing of the two high resolution monitors and the service monitor with a damp cloth. • Wipe the glass with a cloth moistened with a lukewarm aqueous solution of mild detergent. MammoReport Plus Register 8 SPB7-420.831.20 Page 2 of 2 Siemens Medical Solutions Rev.
Page 9
Perform calibration according to the MammoReport System Manual Calibration of Monitors (SPB7- 420.820.20...) • Plus Check calibration according to the MammoReport Quality Control Manual (SPB7- 420.621.20 ...). Siemens Medical Solutions Register 8 SPB7-420.831.20 Page 1 of 2 MammoReport Plus Erlangen, Germany Rev. 01 01/05...
Page 10
3 - 2 Check of image quality This page intentionally left blank. MammoReport Plus Register 8 SPB7-420.831.20 Page 2 of 2 Siemens Medical Solutions Rev. 01 01/05 CSPS 24 Erlangen, Germany...