System Manual
Maintenance Instructions
System maintenance
The Maintenance Protocol
SPB7-420.832.20.01.02 is required for these
Instructions
Print No.: SPB7-420.831.20.01.02
Plus
SP
© Siemens AG 2005
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
_or_ design,_are_
model
reserved.
English
Doc. Gen. Date: 01/05

Advertisement

Table of Contents
loading

Summary of Contents for Siemens MammoReport Plus

  • Page 1 Plus MammoReport System Manual Maintenance Instructions System maintenance © Siemens AG 2005 The reproduction, transmission or The Maintenance Protocol use of this document or its contents is not permitted without express SPB7-420.832.20.01.02 is required for these written authority. Offenders will be liable for damages.
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3: Table Of Contents

    Monitors ......... 2 - 2 3 _______Check of image quality __________________________________________ 3 - 1 4 _______Changes to previous version _____________________________________ 4 - 1 Siemens Medical Solutions SPB7-420.831.20 Page 1 of 2...
  • Page 4 0 - 2 Contents Page MammoReport Plus SPB7-420.831.20 Page 2 of 2 Siemens Medical Solutions Rev. 01 01/05 CSPS 24 Erlangen, Germany...
  • Page 5: Safety Information

    • Maintenance Protocol (SPB7- 420.832.20...) • Plus MammoReport Quality Control Manual (SPB7-420.621.20 ...) Time required Approximately 2 - 2.5 hours for one CSE. Siemens Medical Solutions Register 8 SPB7-420.831.20 Page 1 of 2 MammoReport Plus Erlangen, Germany Rev. 01 01/05...
  • Page 6: Text Emphasis

    Quality Check Image Quality Check System Software Maintenance The work steps in this document that are identified by these abbreviations should be checked off in the Maintenance Protocol (SPB7-420.832.20...). MammoReport Plus Register 8 SPB7-420.831.20 Page 2 of 2 Siemens Medical Solutions Rev.
  • Page 7: Computer

    Listen to the noise of the fans. Does it sound ok? • Shut down the operating system properly and turn off the computer and any external devices. • Disconnect the power supply cable. Siemens Medical Solutions Register 8 SPB7-420.831.20 Page 1 of 2 MammoReport Plus Erlangen, Germany Rev.
  • Page 8: Cables

    Clean the housing of the two high resolution monitors and the service monitor with a damp cloth. • Wipe the glass with a cloth moistened with a lukewarm aqueous solution of mild detergent. MammoReport Plus Register 8 SPB7-420.831.20 Page 2 of 2 Siemens Medical Solutions Rev.
  • Page 9 Perform calibration according to the MammoReport System Manual Calibration of Monitors (SPB7- 420.820.20...) • Plus Check calibration according to the MammoReport Quality Control Manual (SPB7- 420.621.20 ...). Siemens Medical Solutions Register 8 SPB7-420.831.20 Page 1 of 2 MammoReport Plus Erlangen, Germany Rev. 01 01/05...
  • Page 10 3 - 2 Check of image quality This page intentionally left blank. MammoReport Plus Register 8 SPB7-420.831.20 Page 2 of 2 Siemens Medical Solutions Rev. 01 01/05 CSPS 24 Erlangen, Germany...
  • Page 11 Changes to previous version 4 - 1 Changes to previous version Chapter Page Change Initial version Siemens Medical Solutions Register 8 SPB7-420.831.20 Page 1 of 2 MammoReport Plus Erlangen, Germany Rev. 01 01/05 CSPS 24...