MAMMOMAT Novation
Installation Instructions
System
with WH AWS
Print No.:
SPB7-250.812.03.02.02
Replaces: SPB7-250.812.03.01.02
SP
2004
© Siemens AG
The reproduction, transmission or use
of this document or its contents is not
permitted
without
express
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 08.04
written

Advertisement

Table of Contents
loading

Summary of Contents for Siemens MAMMOMAT Novation SP

  • Page 1 W e l s i n g MAMMOMAT Novation Installation Instructions System with WH AWS 2004 © Siemens AG The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages.
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3: Table Of Contents

    Connecting ............33 Siemens AG SPB7-250.812.03.02.02...
  • Page 4 Cleaning ............33 6 _______ Changes since last version_______________________________________ 35 MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 4 of 36 08.04 CS SD 24...
  • Page 5: Configuration

    • the detector and • the syngo Acquisition Workstation. Depending on the installation configuration, some points may be omitted. Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 5 of 36 Medical Solutions 08.04 CS SD 24...
  • Page 6: Tools

    Prerequisites Tools Item Remarks Standard installation and service tools Torque wrench for bolting the stand/console to the floor Electrical screwdriver with adjustable torque is recommended MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 6 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 7: Symbols

    Important Notes Symbols 2Important Notes This symbol indicates exposure of X-ray radiation. This symbol indicates that the measured values have to be entered in the Installation Pro- tocol (SPB7-250.813.01...). Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 7 of 36 Medical Solutions 08.04...
  • Page 8: Factory Adjustments

    The equipment is set to 400 V, 2-phase on delivery. MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 8 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 9: Protective Measures

    Remove the protective strips only when the covers are to be mounted or when vertical adjustment of the swivel arm system is necessary. Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 9 of 36 Medical Solutions 08.04...
  • Page 10: Measurements With The Oscilloscope

    A life-threatening hazard of electric shock exists. For measurements in which a resulting ground loop may falsify the measuring result, use the differential amplifier (difference measurement). MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 10 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 11: Scope Of Delivery

    NOTE The detector package is temperature sensitive, from 10° to 40° Cel- sius. NOTE The detector package has to be disposed locally, after the installa- tion. Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 11 of 36 Medical Solutions 08.04 CS SD 24...
  • Page 12: Unpacking The System Components

    (5/Fig. 1 / p. 12) (two persons are re- quired). 5. Loosen the six bolts(6/Fig. 1 / p. 12) and remove the pallet (5/Fig. 1 / p. 12). MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 12 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 13: Radiation Protection Shield And Control Console

    SPB7-230-031.07.xx.xx. Acquisition workstation 1. Open the cardboard boxes and lift out the computer, monitor and accessories. 2. Assemble the computer according to the user guide that is supplied with the system. Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 13 of 36 Medical Solutions 08.04...
  • Page 14: Removing The Transport Safeguards

    4 / p. 14) and remove the tube head screws(2/Fig. 4 / p. 14) 2. Remove the red transport safeguard (3/Fig. 4 / p. 14). MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 14 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 15: Generator Transportation Safety Plates

    Screws for fixing the generator 1. To be able to access the inside of the stand you have to loosen all marked screws in (Fig. 5 / p. 15)) on the generator panel. Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 15 of 36 Medical Solutions 08.04...
  • Page 16: Lifting Carriage

    The lifting carriage transport safeguard shall be kept for service purposes(Fig. 7 / p. 16) 2. Cut and remove the cable tie securing the balancing spring during transport. MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 16 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 17: Rotary Motion

    The screw and the spacer should be kept for the mounting of the lifting carriage transport safeguard. 2. Remove the transport safeguard on the balancing spring. Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 17 of 36 Medical Solutions 08.04...
  • Page 18: Check Ball Bearings

    (Fig. 9 / p. 18) Check that all twelve ball bearings are in place and that they show no sign of damage. Fig. 9: Ball bearings MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 18 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 19: Protective Strips For The Metal Curtain

    Make sure the protective strips are removed before adjustments are carried out. Holders(2/Fig. 10 / p. 19) for storing the protective strips when not in use are provided on both sides of the curtain. Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 19 of 36 Medical Solutions 08.04...
  • Page 20: Removing The Brick Transportation Screws

    Unpacking the system components Removing the BRICK transportation screws Remove the transportation screws(1/Fig. 11 / p. 20). Fig. 11: Remove BRICK transportation screws Pos. transportation screws MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 20 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 21: Room Planning

    12 / p. 21) • For safety reasons, the acquisition workstation must be set up outside the area, per IEC 60601-1-1(3/Fig. 12 / p. 21) Fig. 12: Room planing Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 21 of 36 Medical Solutions 08.04...
  • Page 22 Other arrangements are possible, provided the neces- sary unit distances based on the available cable lengths are main- tained. For more details on room planning, see the Planning Guide SPB7-250-891.02.xx.xx MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 22 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 23: Installing The Cable Ducts

    4. At the radiation shield, cover the cable entry not used with the enclosed cover plate(6/Fig. 13 / p. 23). Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 23 of 36 Medical Solutions 08.04...
  • Page 24: Aligning The Stand

    This is to ensure optimum stability of the stand. 2. Cover the levelling screws with the cover discs supplied. Fig. 14: Levelling screws Pos. 1 screws MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 24 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 25: Free-Standing Radiation Shield (Option)

    Ensure that the control console is attached to the wall with adequate safety margins. Dow- els and screws are not supplied. Please obtain these locally and ensure that they are suit- able for the material used in the wall. The control console has a weight of 4 kg. Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation...
  • Page 26 (Fig. 19 / p. 27) shows how the control console is to be mounted on the free-standing radiation shield (option), see separate instructions enclosed with the radiation shield. MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 26 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 27 Fig. 19: Control console on free-standing Fig. 18: Control console mounted on table radiation shield (option) Pos. 1 Welded nut Pos. 2 Table top Pos. 3 Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 27 of 36 Medical Solutions 08.04 CS SD 24...
  • Page 28: Bolting The Stand/Radiation Shield To The Floor

    Fig. 20: Base plate for the stand Pos. 1 screws Pos. 2 Cable lead-in from the floor left or right. Obtain cable conduits on site. Pos. 3 Slots for cable lead-in MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 28 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 29: Example Of The Leveling Screws

    22 / p. 30) With installation on the solid floor, cut out the floor covering. Con- crete quality for the solid floor, min. C20/25. Mounting materials must be obtained locally Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 29 of 36 Medical Solutions 08.04...
  • Page 30 The screws must not come in contact with any conductive materi- als such as reinforcing iron, since the grounding could be compro- mised. Stay clear of conductive materials when fastening the screws. MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 30 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 31: Mount The Metal Springs On The Base Plate

    Pos. 3 Metal springs 1. Mount eight metal springs(1/Fig. 23 / p. 31) on the base plate (2/Fig. 24 / p. 31)(3/Fig. 25 / p. 31) Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 31 of 36 Medical Solutions 08.04 CS SD 24...
  • Page 32: Control Console Cable

    Connect the control console cable to the bottom of the MAMMOMAT stand and to the control console. Fig. 26: Control console connection Fig. 27: Location of Control console connection Pos. 1 cable connection MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 32 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 33: Fibre Optics Cable

    The DirectRay Interface card is PCI-bus compatible. Fig. 28: PCI card connection Pos. 1 RX fiber optic cable (shown with cap on conector) Pos. 2 TX fiber optic cable Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 33 of 36 Medical Solutions 08.04...
  • Page 34 BRICK in the MAMMOMAT stand. Fig. 29: BRICK connection Pos. 1 RX 605 fibre optics cable Pos. 2 TX 604 fibre optics cable MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 34 of 36 08.04 CS SD 24 Medical Solutions...
  • Page 35 Page Change DMS conversion. Prerequisites Adapted, so it fits for OPDIMA as well. Scope of deliv- Added a not to dispose the detector package locally. Siemens AG SPB7-250.812.03.02.02 MAMMOMAT Novation Page 35 of 36 Medical Solutions 08.04 CS SD 24...
  • Page 36 Changes since last version MAMMOMAT Novation SPB7-250.812.03.02.02 Siemens AG Page 36 of 36 08.04 CS SD 24 Medical Solutions...