T .O . 33K8-4-551-1
Figure 2.
4 .3 .10 The TI display must indicate within the corresponding values listed in the Limits column of Table 7.
4 .3 .11 Set the Calibrator OPR/STBY switch to STBY.
4 .3 .12 Select the next Resistance Standard (2 .10) as listed in the Resistance Standard column of Table 7.
4 .3 .13 Repeat steps 4 .3 .5 through 4 .3 .12 for each remaining value listed in Table 7 using the appropriate Resistance
Standard (2 .10) .
Table 7.
Range (ADC)
Resistance Standard
Applied (ADC)
Limits (ADC)
100
1.1
1 kS2/Step (10 k12)
100 p .
99 .99707 to 100.00293
µ
1 m
100 SI/Step (1 k12)
1 m
0 .9999737 to 1 .0000263 m
10m
1 IcQ/Step (100 S2) *
10 m
9 .999737 to 10 .000263 m
100 m
100 1-2/Step (10 S2) *
100 m
99 .99593 to 100 .00407 m
1
10 S2/Step (112) *' **
1
0 .9998797 to 1 .0001203
* Connect 10 Resistors in parallel.
** Heat build-up in the 1 S2 shunt [Standard Resistor (2 .10)] will affect accuracy . If test exceeds one minute,
press Calibrator OPR/STBY switch to STBY and wait five minutes before repeating test.
4 .3 .14 Disconnect the test setup.
TI
0
0-
0--
HI 0
DIGITAL
VOLTMETER LO 0
0
0
0
A10
AO
HI
CALIBRATOR
24