HP 3458A Technical Manual page 29

Digital voltmeter
Hide thumbs Also See for 3458A:

Advertisement

T .O . 33K8-4-551-1
Table 8. (Cont)
Range (AC)
Applied (VAC)
Limits (VAC)
10V
10 @20 Hz
9 .99863 to 10.00137
10 @ 1 kHz
9 .99905 to 10 .00095
10@20 kHz
9 .99837 to 10 .00163
10@50 kHz
9 .99676 to 10 .00324
10 @ 100 kHz
9 .99176 to 10 .00824
100 V
100 @20 Hz
99 .9749 to 100 .0251
100 @ 1 kHz
99 .9778 to 100 .0222
100@20 kHz
99 .9778 to 100 .0222
100 @ 50 kHz
99 .9623 to 100 .0377
100 @ 100 kHz
99 .8776 to 100 .1224
1000 V
200 @ 40 Hz
199 .900 to 200 .100
300 @ 10 kHz
299 .800 to 300.200
500 @ 100 kHz
498 .459 to 501 .541
1000 V
(Opt HO1 only)
1k@30 kHz *2
998 .773 to 1001 .227
* I Record the TI indication.
*2 If required, manually range the TI and AC Measurement Standard.
4 .4 .11 Disconnect the test setup.
4 .4 .12 Set the TI to the first range listed in the Range column of Table 9.
4 .4 .13 Connect the Calibrator WIDEBAND output, using the cable and 50 S2 termination supplied with the Calibrator,
to the AC Measurement Standard INPUT 2.
4 .4 .14 Set the AC Measurement Standard to measure the first value listed in the Applied column of Table 9.
4 .4.15 Set the Calibrator for the first value listed in the Applied column of Table 9 and set the OPR/STBY switch to
OPR.
4 .4 .16 Edit the Calibrator controls until the AC Measurement Standard indicates the value listed in the Applied
column of Table 9.
4 .4 .17 Set the Calibrator OPR/STBY switch to STBY.
4 .4 .18 Disconnect the AC Measurement Standard from the Calibrator WIDEBAND output .
27