CXDI-50G
Change of the image read-out method from the sensor was made to the former CXDI-50G. In the
former CXDI-50G, Aux. Shift Register Driver
(PCB-50A) and A/D Conversion (PCB-50AD) are located in the both sides of LANMIT. One panel
is considered as two virtual panels, and image data was read out from both sides. However, in new
CXDI-50G, Aux. Shift Register Driver (PCB-D-EP), Signal Read-out Circuit (PCB-50A) and the
A/D Conversion (PCB-50AD) are located in one side of LANMIT and the image data was read from
one side. Since the appearance of new CXDI-50G is the same as that of the former CXDI-50G, they
are identified by changing the serial number of the main unit.
CXDI System Software
Clinic Software
Serial number of main units
CXDI-50C
Fluorescent substance of CXDI-50G has been changed to CsI from GOS in CXDI-50C so that the
CXDI-50C can be ranked as the more superior performance model (high sensitivity model) than
CXDI-50G. The read-out method of image data is read-out from both sides of LANMIT which is
same as CXDI-50G.
CXDI System Software
1. General
General
(PCB-D-EP), Scanner with Main Scanner Driver
Ver.6.4 and later
Ver.1.04 and later
~
7M 150001
~
3M 250001
Ver.6.4 and later
~
7M 100001
~
3M 200001
1