ABB 9438 User Manual page 28

Low and high level dissolved oxygen
Table of Contents

Advertisement

...7 PROGRAMMING AND ELECTRICAL CALIBRATION
...7.7 Factory Settings Page
– – – – –
Adjust RTX Span
– – – – –
Adjust RTX Zero 2
– – – – –
Adjust RTX Span 2
2
Cal Time 1
(min)
30
Cal Time 2
(min)
0 0 0 0 0
Alter Fact. Code
1 8 . 1
D.O.
26
Adjust Retransmission Span
Set the milliammeter reading to 20.00 mA.
Note. Retransmission signal span is calibrated using 20.00 mA. The correct value
transmitted depends on the range selected in the
Adjust Retransmission Zero 2
See
Adjust Retransmission Zero
Adjust Retransmission Span 2
See
Adjust Retransmission Span
Calibration Time 1
Wait period before the stability of the sensor is checked during the calibration sequence.
Programmable from 1 to 10 minutes (default = 2 minutes). See Section 7.5.
Calibration Time 2
Recovery period where the sample is allowed to flow and the instrument settles on
reading, before the instrument is brought back on-line (default = 30 minutes).
See Section 7.5.
Alter Factory Setting Security Code
Set the security code to a value between 00000 and 19999.
Return to
Operating Page
g/kg
Set Up Outputs Page
.
.
, Section 6.2.1.
.