Performance
Measured value error (under reference conditions)
• Measuring error
• Dead zone
• Additional contribution of
analog output
• Influence of pressure
(air, 20°C)
• Long-term stability
• Repetitive accuracy
Absolute Measuring Error
Mechanical
Flange
• Process Connection
• Materials of the wetted
parts – in contact with
the process
• Pressure (vessel)
1.
Flange material may be either 316/316L or 1.4571 at the discretion of Siemens
Milltronics Process Instruments Inc. Actual flange material will be noted on the
side of the flange.
7ML19985FH03
≤ ± 15 mm at 0.26 to 2 m distance
≤ ± 5 mm at 2 to 10 m distance
≤ ± 15 mm at 10 to 45 m distance
(see next page)
0-260 mm from bottom edge of flange
≤ 0.1% of the measured value
≤ 0.05% of full scale
≤ 0.3% to 10 bar without pressure correction
≤ 2 % at 10 to 64 bar without pressure correction
≤ ± 1 mm/year
≤ ± 1 mm at 0 to 45 m, damping ≥ 1 s
20
15
10
5
0
-5
-10
-15
-20
2
10
Distance from flange [m]
Flange DIN 2527, ANSI B16.5, or JIS B2238 equivalent bolt
pattern
Stainless steel
emitter (or glass/PTFE, Zone 0 and Zone 20 devices)
Pressure rating of flange is dependent on temperature.
Refer to Appendix IV for charts, or obtain the reference
drawing listed on the flange device tag.
SITRANS LR 400 – INSTRUCTION MANUAL
20
30
1
flange and CF8M horn, PTFE
40
45
Page 5