7 Special functions of the DMS Go+ TC
– Apply some couplant to the reference block at the in-
strument rear and to the reference block made of
copper.
– In the function group PROBE&CAL, select the func-
tion PROBE ZERO.
– Press one of the left or right arrow keys in order to
start the zeroing. The message Put probe on zero
block - use couplant is displayed on the screen.
– Couple the probe. The calibration starts.
– Keep the probe coupled as long as the message Ac-
quiring, please stay coupled is visible on the dis-
play screen. When the message is no longer visible,
the calibration for the measurement mode Auto-V is
finished.
You can then carry out wall thickness measurements on
non-coated test objects. Every time you couple the
probe, you can see the current sound velocity in the test
object at the top of the display screen.