Guidance and manufacturer's declaration-electromagnetic immunity for all EQUIPMENT
The LCD display for Healthcare Applications is intended for use in the electromagnetic
environment specified below. The user of the LCD display for Healthcare Applications should
assure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
3 Vrms
Conducted RF
3 V/m
Radiated RF
80 MHz to 2.5
IEC 61000-4-3
GHz
12
AND SYSTEMS that are not LIFE-SUPPORTING
Compliance
Level
3 Vrms
3 Vrms
Electromagnetic Environment-Guidelines
Portable and mobile RF communications
equipment should be used no closer to any part
of the LCD display for Healthcare Applications
and should assure that it is used in such an
environment,
including
recommended separation distance calculated
from the equation applicable to the frequency of
the
transmitter.
distance
80MHz to 800 MHz
800 MHz to 2.5GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres(m)
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey*
should be less than the compliance level in
each frequency range*
Interference may occur in the vicinity of
equipment marked with the following symbol:
cables,
than
Recommended
separation
2
.
the
1
,