Immunity
Electromagnetic immunity
The InnoSpire Go is intended for use in the electromagnetic environment specified below. The customer or
the user of the InnoSpire Go should assure that it is used in such an environment.
Immunity test
Compliance level
Electrostatic
±8 kV contact
discharge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power supply lines
transient/burst
IEC 61000-4-4
Surge
± 1 kV differential mode
IEC 61000-4-5
Voltage dips, short
0% UT; 0.5 cycle at 0°, 45°,
interruptions and
90°, 135°, 180°, 225°, 270°
voltage variations
and 315°
on power supply
0% UT; 1 cycle
input lines
70% UT; 25/30 cycles
IEC61000-4-11
0% UT: 250/300 cycles
Single phase: at 0°
Power frequency
30 A/m
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
54
Electromagnetic environment – guidance
The relative humidity should be at least 5 %.
Mains power quality should be that of a typical
home commercial or hospital environment.
Mains power quality should be that of a typical
home commercial or hospital environment.
Mains power quality should be that of a typical
home commercial or hospital environment. If
the user of the InnoSpire Go requires continued
operation during power mains interruptions
beyond that provided by the battery, it is
recommended that the InnoSpire Go is powered
from an uninterruptible power supply.
Power frequency magnetic fields from common
appliances in the home are not expected to affect
the device.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment. Keep the
InnoSpire Go away from sources of high levels of
power line magnetic fields (in excess of 30 A/m)
to reduce the likelihood of interference.
Immunity
Electromagnetic immunity
The InnoSpire Go is intended for use in the electromagnetic environment specified below. The customer or
the user of the pump should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Conducted RF
3 Vrms
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
6 Vrms in ISM
and amateur
radio bands
Radiated RF
3 V/m
10 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
ENGLISH
Electromagnetic environment – guidance
The InnoSpire Go is suitable for the
electromagnetic environment of typical
homes, commercial or hospital settings.
55