After a successful leak test, set the instrument to indicator mode:
1. Connect the pressure standard for the pressure range to be checked.
2. Set the pressure standard to zero pressure, if necessary press the zero key
so that the instrument display shows zero.
3. Adjust the pressure standard to the 0% span pressure value.
4. Compare the pressure value of the pressure standard to the value displayed
and record the difference.
5. Repeat steps 3 and 4 for the 100% span pressure value.
6. If the recorded difference exceeds the tolerance in the table, the instrument
needs a calibration adjustment.
7. Adjust the pressure standard to zero pressure.
The DPI 150 instrument has the following pressure ranges.
Transducer
range
bar
0.025
0.07
0.2
0.35
0.7
1
2
3.5
7
10
20
35
70
100
135
200
350
700
K382 Issue No. 1
Applied Pressure
% span
0
100
Equivalent Pressure Units
mbar
psi
25
036
70
1
200
3
350
5
700
10
1000
15
2000
30
3500
50
7000
100
10000
150
20000
300
35000
500
70000
1000
N/A
1500
N/A
2000
N/A
3000
N/A
5000
N/A
10000
Tolerance
± 0.01%FS
± 0.01%FS
Table 4
Pa
hPa
2500
25.0
7000
70
20000
200
35000
350
70000
700
100000
1000
200000
2000
350000
3500
700000
7000
1000000
10000
2000000
20000
3500000
35000
7000000
70000
N/A
100000
N/A
135000
N/A
200000
N/A
350000
N/A
700000
6
kPa
MPa
2.5
0.0025
7
0.007
20
0.02
35
0.035
70
0.07
100
0.1
200
0.2
350
0.35
700
0.7
1000
1
2000
2
3500
3.5
7000
7
10000
10
13500
13.5
20000
20
35000
35
70000
70

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Druck dpi 142