Power Distribution Board; Ambient Conditions - Toshiba Activion 16 Product Data

Multislice helical ct scanner
Table of Contents

Advertisement

Power distribution board

In case of
In case of
3-phase, 400 V
3-phase, 200 V
NFB
150 A
System

Ambient conditions

Temperature
Scan room
Gantry
20°C to 26°C
Tolerance: ±2°C No
Patient couch 20°C to 26°C
Tolerance: ±2°C No
Operator's room
Console
16°C to 28°C
*1: When scanning is not performed.
*2: When scanning is performed continuously at the maximum
rated output of the system.
Minimum area for installation
• For the long patient couch version:
• For the short patient couch version:
10
Ground resistance:
As per applicable
legal requirements.
NFB
100 A
Ground bar
System
transformer
(option)
Humidity
Heat generation
40% to 80%
Approx.
9.720 kJ/h (*1)
condensation 32,070 kJ/h (*2)
40% to 80%
Approx.
1,080 kJ/h (*1)
condensation 1,800 kJ/h (*2)
40% to 80%
Approx.
No
9,720 kJ/h
condensation
2
22 m
2
21 m
Room layout example
3,072
Gantry
1
Patient couch
2
*: For the short patient couch version
1
2
3
Unit: mm
Console
3