Performance Specifications - Toshiba Activion 16 Product Data

Multislice helical ct scanner
Table of Contents

Advertisement

SURE
TM
Fluoro
(option)
Conventional CT fluoroscopy shows only a single slice,
SURE
but
Fluoro (Multislice CT fluoroscopy) permits realtime
image reconstruction to display 3 images obtained by
combining data from the SSMD.
improves operability in biopsy and interventional proce-
dures.
• Tilt Helical scanning
Helical scanning with gantry tilt, from 30° forward to 30°
backward, is available because the system employs the
TCOT reconstruction technique.
COMPOSITION
Standard composition (Model: TSX-031A)
• Gantry ................................................................................1
• Patient couch......................................................................1
• Console ........................................................................1 set
• Accessories
– Inter-unit cables
– Manuals
– Set of phantoms
– Acquisition support
– Footswitch for the patient couch
Note: The console desk is not included in the standard con-
figuration.
Optional items
• Cerebral blood-flow analysis system
(CBP-study) (CSCP-002A)
• System transformer (CETF-004B)
• Display system for dental application (CDP-07A)*
• FlyThrough software (CFT-03A)*
• Vessel view (CVV-001A)
SURE
Fluoro (TSXF-003E)
SURE
• LCD monitor for
Fluoro (15-inch type) (CMM-003E)
• Magneto-optical disk drive unit (CMO-12A)
• DICOM storage SCP (COT-30D)
• DICOM MWM (COT-32D)
• DICOM MPPS (COT-33D)
• DICOM Q/R SCP (COT-34D)
• DICOM Q/R SCU (COT-35D)
• DICOM storage commitment SCU (COT-41D)
• DICOM PGP profile (COT-44A)
• DICOM fast transfer system (COT-45A)
• Color printer interface (CCP-03A)
• ECG-gated scan system (CHEG-004C)
• Injector synchronization system (CKIS-004A)
• Orbital synchronized scan system (CKOS-001A)*
• Pediatric scanning system (CHKS-002A)
* Not available in the U.S.A.
2
SURE
Fluoro significantly

PERFORMANCE SPECIFICATIONS

Scan parameters
• Scan regions:
Whole body, including head
• Scan system:
360° continuous rotate/rotate
• Scan plan
programming:
More than 360 different
sequences can be pre-pro-
grammed.
• Scan time
– CT scan:
0.48 s (Partial), 0.75, 1, 1.5, 2,
and 3 s
• Scan cycle time (for 0.75-s scan)
– SCAN & SCAN mode: Min. 1.5 s (rapid sequence
scanning, couch-top movement
10 mm)
Note: The scan cycle time refers to the time between the start of
one scan and the next.
SCAN & VIEW mode permits immediate viewing of
images after acquisition of each individual slice.
• Scan field
φ180 mm (SS)
– CT scan:
φ240 mm (S)
φ320 mm (M)
φ390 mm (L)
φ500 mm (LL)
– Scanoscopy:
Axial direction
Up to 390 mm
200 mm to 1,750 mm (1,450 mm*)
*: For the short patient couch version
Note: The actual range that can be viewed is less than the
couch-top movement range in scanoscopy.
• Slice thickness:
0.5, 1, 2, 3, 4, and 5 mm
These slice thicknesses are implemented by stacking the
data acquired in one of the following acquisition modes.
• Acquisition
0.5 mm × 16 rows
– 16-row
1 mm × 16 rows
0.5 mm × 4 rows
– 4-row
1 mm × 4 rows
2 mm × 4 rows
3 mm × 4 rows
4 mm × 4 rows
5 mm × 4 rows
• Gantry tilt angle:
From forward 30° to backward
30° (in 0.5° increments)
Remote control from the console
is possible.
• Tube position for
scanoscopy:
0°, 90°, 180°, and 270°
Any arbitrary angle can be
specified (in 5° increments).
• Gantry aperture:
720 mm in diameter
Longitudinal
direction
Adjustable from

Hide quick links: