Philips System 21 Quick Operating Manual page 18

Instrument bus system
Table of Contents

Advertisement

Function I Command
Description
Default
See
Service Request masking
D
"MSR
n"
Mask Service Request with
"MSR
0"
PM2101
decimal value
n=
0 .... 511
Mask-values:
values may be added to enable
more conditions to generate SRQ.
decimal
enable Service Request when:
value
256
PM2101 is no longer BUSY
128
A System 21 event occurred
16
PM2101 received an illegal code
0
"MSK"+ <digits>
Mask System 21 events
"MSK
0"
Sys.M
(maximum 9
digits).
A zero digit
disables the corresponding event
to generate a Service Request.
digits:
"1"
A slave had a power fail,
"0"
received a power-on reset or
received an illegal code or
illegal
sequence.
"
2"
An addressed module is not
present.
"0"
"3"
All units became ready
performing the programmed
"0"
actions.
''4''
All units are ready with the
"0"
received
data and may receive
new data.
"5"
A unit has
data
available.
"0"
"6"
All units are ready with the
"0"
block or
scan.
"7"
A unit has a warning.
"0"
"8"
The ready-line became high
(ready true)
"0"
"9"
A trigger-pulse was captured on
"0"
the trigger-line.
18

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents