Transmitter Specification - ABB AZ10 Operating Instruction

Combustion gas analysis.
Hide thumbs Also See for AZ10:
Table of Contents

Advertisement

A Z 1 0 O X YG E N A N A LY Z E R | C O M B U S T I O N G A S A N A LY S I S | O I/A Z 1 0 - E N R E V. A

12 Transmitter specification

Transmitter enclosures
Remote
Wall-mounted
4 gland entries
Optional 1/2 in NPT, M20
Automatic calibration
AutoCal hardware
• Isolated solenoid valve control as standard,
24 V @ 2 W per valve*
• Dedicated isolated digital inputs to monitor pressure
switch contacts as standard – voltage-free, normally
closed with gas present
Display and switches
Display type
Graphical 128 x 64 pixel LCD
Display backlight
Green LED
Operator switches
4 capacitive switches (operated through the front glass)
Relay outputs
Number
2 standard
Type
Normally closed, 5 A @ 230 V AC
or 30 V DC (non-inductive)
Functions
User-configurable – can be activated by one or more of the
following signals:
• Process alarm 1, 2, 3, 4
• Calibration in progress
• Calibration failed
• Out of test gas 1, 2
• Test gas 1 valve control
• Test gas 2 valve control
• Failure diagnostic
• Out-of-specification diagnostic
• Maintenance required diagnostic
• Function check diagnostic
*For driving internal automatic calibration (AutoCal) probes or can be used
to drive external calibration units on remote transmitters only.
Analog outputs
Standard
• 1 isolated current output
• Programmable to retransmit oxygen (linear or logarithmic)
or temperature
• Programmable over 4 to 20 mA
• Over-range capability to indicate system failure
programmable from 4 to 22 mA
Optional
• 1 isolated current output
• Programmable to retransmit oxygen (linear or logarithmic)
or temperature
• Programmable over 0 to 20 mA
• Over-range capability to indicate system failure
programmable from 0 to 22 mA
31