Milltronics ILE-37 Sensing Head Diaphragm Replacement
Objective:
The procedure for replacing the diaphragm on a Milltronics ILE-37 sensing head.
•
Equipment:
•
•
While every effort was made to verify the following information, no warranty of accuracy or usability is expressed or implied.
Overview:
To replace the diaphragm on a Milltronics ILE-37, please follow the directions below.
For an ILE-37 that does not have an LVDT conditioner card, skip
Part One: Remove Diaphragm
Step One: Remove Fiberglass Cover
1.
Use 6 mm hex screwdriver or Allen key to loosen the
four bolts connecting the fiberglass cover to the
main body.
Siemens Milltronics Process Instruments Inc.
1954 Technology Drive, P.O. Box 4225, Peterborough, ON Canada K9J 7B1
Tel: (705) 745-2431 Fax: (705) 741-0466
Email: techpubs.smpi@siemens.com
Milltronics ILE-37
Replacement diaphragm
1 ¾" or 45 mm wrench, or adjustable
wrench
www.siemens.com/processautomation
APPLICATION GUIDE
•
10 mm wrench
•
6 mm hex screwdriver or Allen key
•
5 mm Allen key
•
2.5 mm hex screwdriver or Allen key
Part One: Step Four
Part Two: Step Three
and
2.
Remove the four bolts and washers.
3.
Remove fiberglass cover.