Installing

3.1

Basic safety instructions

Hot surfaces resulting from hot process media
Danger of burns resulting from surface temperatures above 70 °C (155 °F).
● Take appropriate protective measures, for example contact protection.
● Make sure that the maximum ambient temperature is not exceeded by implementing
Wetted parts unsuitable for the process medium
Danger of injury or damage to device.
Hot, toxic and corrosive media could be released if the process medium is unsuitable for the
wetted parts.
● Ensure that the material of the device parts wetted by the process medium is suitable for
Note
Material compatibility
Siemens can provide you with support concerning selection of analyzer components wetted
by the process medium. However, you are responsible for the selection of components.
Siemens accepts no liability for faults or failures resulting from incompatible materials.
Unsuitable connection parts
Danger of injury or poisoning.
In case of improper mounting hot, toxic and corrosive process media could be released at
the connections.
● Ensure that connecting parts (such as flange seals and screws) are suitable for the
Devices in explosion-proof design
Compact Operating Instructions, 05/2015, A5E35134047-02
CAUTION
protective measures. Refer to the information in the chapter "Technical specifications
(Page 79)".
WARNING
the medium. Refer to the information in the chapter "Technical specifications (Page 79)".
WARNING
connection and the process media.