ABB Z-MT Commissioning And Calibration Procedures page 15

Table of Contents

Advertisement

Cell temp
T/C
or
set
or
T/C
or
Enter
Cell T/C type K
or
R
S
K
or
Enter
ANALOGUE RETRANSMISS
Twice
OP1 ReTx xxxxx
or
or
OP1 ReTx Cell T
or
Enter
Vary Lo OP1
& ENT
or
Enter
Vary Hi OP1
& ENT
or
Continued from opposite page.
Cell Temperature, Thermocouple or Preset Temperature
Select thermocouple for cell temperature measurement.
Store.
Advance to next parameter.
Cell Temperature Thermocouple Type
Select the thermocouple being used for cell temperature measurement:
Make a note of the thermocouple type and then reset it to Type K.
Store.
Advance to next parameter.
Select the Analogue Retransmission P .
Operate for approximately three seconds to gain access to the SECURE
PARAMETERS (retransmission 1)
Advance to SECURE PARAMETERS (retransmission .
SECURE PARAMETERS (retransmissio
Retransmission Output 1
Make a note of the parameter range assigned to retransmission output 1 and
then assign the cell temperature range.
Store.
Advance to next parameter.
Retransmission Ouput 1, Lower Retransmission
Make a note of the value on the upper display and then reset it to '0.0'.
Store.
Advance to next parameter.
Retransmission Ouput 1, Higher Retransmission
Make a note of the value on the upper display and then reset it to '1400'.
Continued overleaf.
within the Analogue Retransmission Page.