17
- Set potentiometer R 18 to the centre position.
- Tum the spindle of double potentiometer RI until the nominal fre­
quency
is obtained.
- Set the frequency dial to position 15 and tighten the nut without moving
the spindle of double potentiometer RI.
- Check the frequency di al setting at 1500
Hz.
2. Electrical check
- Set the frequency dial to position 100.
- Set SK3 to position ""'.
- Set SK2 to position
x
100.
- Check that the frequency
is
10 kHz
±
2%
(with the beat frequency
method).
- Set the frequency dial to position
10.
- Check that the frequency
is
1kHz
±
2%. If necessary, adjust R18 so
that the frequency inaccuracy is smaller than
2%.
- Set SK2 to position
x
l k.
- Set the frequency dia1 to position 10.
- Check that the frequency is 10 kHz
±
2%
. If necessary, correct the
frequency with C22.
- Set the frequency dial to position 100.
-
Check that the frequency is 100 kHz
±
2%.
If necessary, correct the
frequency with C1.
- Check the frequency at all ranges in the positions
10,
25, 40, 70 and
100 of the frequency dia!. Tolerance:
±
3%
±
0.7 Hz.
C. DISTORTION OF THE SINEWAVE VOLTAGE
- Set SK3 to position ""'.
-
Measure the distortion with a non-linear distortion meter.
The distortion should be smaller than
0.7%
at 20 Hz and 100 KHz.
At l kHz the distortion should be smaller than 0.3%.
D. AMPLITUDE
-
Set attenuator R2 to position 2.0 V.
- Set SK3 to position ""'.
- Check that the output amplitud e (no 10ad at BU 1-BU2) at all fre­
quencies is 2 V
rms.
If necessary, correct with R26.
- Set SK3 to position
'U
.