® Health o meter Professional 2595KL ® Thank you for your purchase of this Health o meter Professional product. Please read this manual carefully, and keep it for easy reference or training. TABLE OF CONTENTS Caution and Warning....................3 Specifications for This Scale ..................4 Certifications / Connectivity Disposal / Service ............
® The Health o meter Professional 2595KL scales were intended to be used in a professional medical environment by trained medical staff. This product was designed to weigh patients that are sitting on the chair. Do not modify the product or use it for anything other than its intended purpose.
GENERAL ® The Health o meter Professional 2595KL Digital Chair Scale uses highly sophisticated microprocessor technology. This precision instrument is designed to provide accurate, reliable and repeatable weight measurements by providing features that make the weighing process simple, fast and convenient. Weight is displayed in kilograms or pounds. The scale includes a seat with arm supports, and a footstool that can be folded up for easy entry/exit.
Customer Service Information The 2595KL Digital Chair Scale is shipped assembled in one carton, except for final installation of scale casters and brakes. Carefully inspect the carton for shipping damage before unpacking. If damage is found, contact Customer Service.
2595KL ASSEMBLY INSTRUCTIONS To avoid damaging the 2595KL when unpacking it, do not use a box cutter, knife, scissors or any sharp object when opening the protective inner packaging. If damage is found, ® contact your shipper or a Health o meter Professional representative immediately.
Page 7
2595KL ASSEMBLY INSTRUCTIONS Assembly Instructions (Refer to Figures 1- 3) Note: Do not attach the AC adapter at this point. Wait until Step 9. Caution: It is recommended that the assembly be performed by two persons. Remove the chair scale assembly from the carton and place chair scale on a flat, level and dry surface.
Page 8
2595KL ASSEMBLY INSTRUCTIONS Figure 2. Assembly of the 2595KL Caster Brake 3/16” (5mm) Allen Wrench edtBCA Figure 3 P/N UM2595KL_REV20200128...
SET UP INFORMATION / BATTERIES Set Up Information (Refer to Figures 4- 5) Place 6 C-cell batteries in the battery holder (see “Replacing Batteries” below), or plug the AC adapter into the chair scale assembly and into the power source. Press the ON/OFF button to turn the scale on.
Page 10
SET UP INFORMATION / BATTERIES Figure 4. Replacing Batteries tdBBCA trloaCA edboC Figure 5 P/N UM2595KL_REV20200128...
OPERATING INSTRUCTIONS Figure 6. Control Panel ITEM DESCRIPTION FUNCTION ON/OFF button Turns scale ON and OFF. REWEIGH button Allows repeated reweighing of person. ZERO button Zeros digital display prior to weighing. Enables the cancellation (tare) of weight if some object was placed on the scale.
MAINTENANCE GENERAL This section provides instructions for maintenance, cleaning, and troubleshooting for the 2595KL Digital Chair Scale. Maintenance operations other than those described in this section should be performed by qualified service personnel. MAINTENANCE Before first use and after periods of non-use, check the scale for proper operation and function.
TROUBLESHOOTING Refer to the following instructions to check and correct any failure before contacting service personnel. SYMPTOM POSSIBLE CAUSE CORRECTIVE ACTION Scale does not turn on Defective battery Replace batteries Faulty electrical outlet Use a different outlet Defective power supply Replace AC adapter Questionable weight or External object interfering...
CALIBRATION PROCEDURE This scale has been factory calibrated, and does not require calibration prior to use. If required, the scale can be calibrated. The following calibration assumes the scale is set to pounds, and 200 lb is employed as the calibration standard. Note: Only weights certified and traceable to national standards are to be used for calibration procedures.