Starting the Sample Gas

Accessing the User
Program
3-4
Initiate the flow of sample gas to the transmitter and allow a few
minutes for the system to reach equilibrium. If the sample gas has a
hydrogen concentration higher than the Low Low alarm setting, the
Low Low alarm resets, and the following data screen appears.
1 H2/AIR
91.52
Note: Allow approximately 30 minutes for the sensor to reach
temperature equilibrium, to ensure accurate readings
Notice that the display shows the slot number and the input range on
the top line, and the current live value along with the unit type are
displayed in the center of the screen.
Note: The "Slot Number" refers to the location of the applicable
circuitry in the digital controller module. Slot 0 is the main
circuit board, slot 1 holds the XDP circuit board, and slots 2-4
may be used for optional circuit boards.
The XDP software permits the operator to configure the meter for his
specific requirements. To accomplish this, it is necessary to leave the
measurement mode and enter the User Program as follows (see the
menu maps in Figure C-1 on page C-1 and Figure C-2 on page C-2):
1 H2/AIR
91.52
Enter the operator level password (2719) at the next screen. This
password grants access only to those menus needed by the operator of
the meter. A different password is required to access the more
advanced menus (see Chapter 5, Advanced Programming, for details).
Password
Enter Password
XXXX
s
This is a typical display. The
actual concentration of your
pcnt
sample gas is shown here.
While in measurement mode,
[ESC]
press the
pcnt
[ESC]
keys in sequence.
Use [ ] and [ ] to position the
cursor under the desired char-
acter. Then, use [ ] and [ ] to
increment the value. Press
[ENTER]
when done.
[ENTER]
,
, and
Operation