Starting The Recording; Auto-Start - GE SEER 1000 Operating Manual

Ecg recorder and mobile application
Table of Contents

Advertisement

Starting the Recording

Press the Event Button in the middle of the recorder to start the record-
ing. The recording only starts if no low battery and no open lead are de-
tected.
CAUTION
SIGNAL QUALITY – If there is a severe muscle interference or artifact
on the ECG when you tap the top of the electrodes lightly or when the
patient moves, it is possible that the electrodes are not attached
properly.
Reattach the electrodes properly.
NOTICE
RECORDING DURATION MAY NOT BE REACHED – In order to achieve
the full 7 day recording duration, a new or fully charged battery is re-
quired. A low-end, non-brand named alkaline battery or a rechargea-
ble battery which is not fully charged or aged may not last 7 days.
Before every recording, the Event Button LED must be green. If it
flashes yellow once every second, a new battery is necessary.
NOTE
If a previous recording is still stored in the recorder, the Event Button
LED will be yellow and the recording will not start. You will need to
erase the recording before continuing. A recording can only be de-
leted using the app or the Holter analysis software.
NOTE
If available, you can also start the recording with the app after select-
ing or adding a patient and after checking the signal quality.
See chapter 5, "Applications (Apps) – General Information", chapter 6,
"Using the Apple iOS Apps", and chapter 7, "Using the Personal Com-
puter App", for more information on the apps.

Auto-Start

If the start sequence has not been completed and the recorder detects a
signal from at least one channel, it automatically starts the recording af-
ter eight minutes.
The eight-minute timer is retriggered
2067634-077 Revision B