Summary of Contents for Siemens Dräger Medical Carina
Page 1
Instructions for Use Carina Sub-Acute Care Ventilator WARNING Software 3.n For a full understanding of the per- formance characteristics of this medical device, the user should carefully read these Instructions for Use before use of the medical device.
Working with these Instructions for Use In the header – the subject of the main chapter To help you find your way around quickly. Operation On each page – the instructions for use Combining text with illustrations. The information is Setting ventilation modes Switching on, Carina shows the ventilation mode last set along with the set parameters.
Trademarks – AutoFlow ® – Carina ® – SyncPlus ® are registered trademarks of Dräger. – BIPAP* – Virex ® is a registred trademark of JohnsonDiversery. Definitions WARNING A WARNING statement provides important information about a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Page 4
This page intentionally left blank Instructions for Use Carina SW 3.n...
For Your Safety and That of Your Patients For Your Safety and That of Your Patients Strictly follow these Instructions for Use ..8 Maintenance ......8 Accessories .
For Your Safety and That of Your Patients Strictly follow these Instructions for Use Safe connection with other electrical equipment WARNING WARNING Any use of the medical device requires full Electrical connections to equipment which is understanding and strict observation of all not listed in these Instructions for Use should portions of these instructions.
For Your Safety and That of Your Patients Functional safety The essential performance consists in controlled and monitored patient ventilation with user-defined settings for the monitoring functions – minimum inspiratory breathing gas flow, – maximum airway pressure, or, if one of the limits set is exceeded, by an appro- priate alarm.
For Your Safety and That of Your Patients General WARNINGS and CAUTIONS WARNING Do not use the device during magnetic reso- The following WARNINGS and CAUTIONS apply to nance (MRI, NMR, NMI)! Device operation may general operation of the device. WARNINGS and be affected, thus placing the patient at risk.
For Your Safety and That of Your Patients CAUTION WARNING Only use the device in well ventilated rooms to In the case of patients who require a high O prevent over heating of the device. concentration ensure that there is an emer- gency supply of oxygen available, e.g.
For Your Safety and That of Your Patients Back-up ventilation with an independent manual ventilation device WARNING If a fault is detected in Carina so that its life- support functions are no longer assured, ven- tilation using an independent ventilation device must be started without delay –...
Intended Use Intended Use Intended Use ......14 Sub-acute care definition ....14 Instructions for Use Carina SW 3.n...
Page 14
Intended Use Intended Use Carina – long-term ventilator for ventilator-depend- ent patients or ventilator-assisted patients. – For treatment of sub-acute care patients in hos- pital or medical rooms. – For invasive ventilation or non-invasive ventila- tion. – For patients with a tidal volume of at least 100 mL.
System Overview Main functions Ventilation modes Ramp Variable rise in pressure from the lower pressure VC-SIMV AF level to the upper pressure level. Volume Control – Synchronized Intermittent Man- datory Ventilation – AutoFlow Intermittent triggered mechanical ventilation (vol- Volume Guarantee ume-controlled) in conjunction with spontaneous VG ensures that patients receive a defined tidal vol- breathing.
System Overview The following are displayed as a bar graph: To protect the patient and the user against electrical – Airway pressure hazards, it is essential that all systems consisting of medical devices as well as other electrical devices Display of: which are not restricted to computers, printers etc., –...
System Overview Control panel, top 1 Carrying handle 2 Screen for application-specific information on ventilation 3 Buttons for selecting functions and ventilation parameters 4 Centre rotary knob for setting and confirming functions and parameters 5 »Audio paused 2 min.« button for suppressing the alarm tone for 2 minutes and for visual indication of alarm status: red = warning, yellow = caution...
System Overview Front connection block 1 Ventilation hose connection 2 Emergency air inlet and oxygen overflow 3 Connection for control hose (for hose system with expiratory valve) Rear connection block 4 Not assigned 5 Connection for nurse call system 4 5 6 7 6 RS 232 serial port for data communication (MEDIBUS) 7 Not assigned...
System Overview Trolley 1-63 cm, Trolley 1-78 cm 1 Carina bracket 2 Standard rail grip 3 Universal bracket with standard rail, optional 4 Trolley column 5 Pedastal 6 Red locking lever of the column mount closed (under pedestal) 7 Dual rollers with locking brakes, 4 each 8 Battery holder, optional Instructions for Use Carina SW 3.n...
System Overview Symbols Mains switch ON Menu 1 Menu 2 Mains switch OFF Menu 3 Connection for nurse call system Observe the Instructions for Use! Data port Gas inlet (emergency air inlet) Power supply display Gas outlet (oxygen overflow) Standby mode ON/OFF Disposal information Turn off audible alarm Connection for control hose...
System Overview Abbreviations Abbreviation Explanation Abbreviation Explanation Apn. Vent. Apnoea ventilation Leakage minute volume leak AutoFlow Automatic optimisation of inspi- Non-invasive ventilation ration flow (mask ventilation) Breaths per minute Pascal, unit of pressure Ventilation strokes per minute 1 mbar = 100 Pa = 1 hPa 1 bar = 100 kPa = 1000 hPa BTPS Body Temperature, Pressure,...
Page 23
System Overview Abbreviation Explanation Delay time for »Airway pressure disconn low !!!« alarm Expiration time Inspiration time setting Ti max Maximum inspiration time setting Tube Tube ventilation (invasive ventilation) UMDNS Universal Medical Device Nomenclature System Nomenclature for medical devices VC-SIMV Volume Control –...
Page 24
This page intentionally left blank Instructions for Use Carina SW 3.n...
Operating concept Control panel Centre rotary knob 1 For setting and confirming functions and venti- lation parameters. To set the value, turn rotary knob. To confirm, press rotary knob. Buttons with a fixed function 2 »Audio paused 2 min.« button for switching off the alarm tone for 2 minutes.
Operating concept Screen Layout 1 Alarm line for indicating audible alarm switched off and alarm message 2 Status line for current device settings 3 Display for real-time waveforms of flow and pressure or bar display for pressure and 4 measured values 4 Display for 2 measured values, configurable 5 Information line, e.g.
Operating concept Display of relevant ventilation parameters 1 When setting a ventilation parameter the rele- vant ventilation values calculated by Carina are indicated in the display field for »Value 1«. When the setting has been confirmed, the dis- play field closes. Colour concept Yellow Function/parameter selected, not...
Operating concept Calling up menus 1 Press »Select Menu« button. All the menus are shown. 2 Press the button for the relevant menu. The functions and parameters of the menu selected are displayed, e.g. »Vent.Set« menu. Expanding menus 3 In the case of menus with more than 4 or 8 func- tions/parameters the other functions/parame- ters can be activated by pressing the button.
Page 30
This page intentionally left blank Instructions for Use Carina SW 3.n...
Preparation Preparation Information on preparation ....32 Connecting to the nurse call system..52 Fitting the inlet filter ..... 32 MEDIBUS interface .
Preparation Information on preparation Before every use, reprocess Carina and all the accessories according to the information in the Instructions for Use, see page 103. Comply with the hospital's hygiene regulations! Fitting the inlet filter Unpack the inlet filter. 1 Insert the inlet filter at the bottom, 2 swivel it in and 3 press at the top until its snaps into place.
Preparation Stability To prevent the device from tipping over, only use Carina up to a maximum inclination of 15°! Carina with trolley Use Dräger trolley "Trolley 1-63 cm" or "Trolley 1-78 cm" only. Before use, check readiness for operation of the trolley: 1 Red locking lever of the column mount closed, 2 Attachments such as standard rail grip or...
Preparation Position Carina on the trolley Open both locks. WARNING Only touch the grip part of the locks. Risk of trapping fingers at the lock! 1 Red areas must be visible. Hold Carina with both hands and place the holes in the underside of the device over the mounting pins.
Preparation Attach universal bracket with standard rail Attach right side of holder to right side of column. 1 Make sure the nose of the universal bracket is completely behind the revolving edge. 2 Attach the left side of the holder to the left side of the column.
Preparation Attaching battery holder 1 Press the locks inwards to open. 2 Open the door of the battery holder. Position the battery holder against the column. Attach the battery holder. Use the four provided bolts. Tighten the four bolts. Use the provided Allen key.
Preparation Use in the hospital WARNING For greater stability, move accessories into their optimal position for moving the trolley: Move hinged arm to its smallest extension. Keep hoses as close to the trolley as pos- sible. Hang the humidifier on the trolley, do not attach to the device! Otherwise there is a risk of falling.
Page 38
Preparation Moving the trolley 1 Release all four locking brakes. 2 Hold and move the trolley by gripping the stand- ard rail or the trolley column. Parking the trolley 3 Engage all four locking brakes. Instructions for Use Carina SW 3.n...
Preparation Information on using bacterial filters, HME systems and hose systems Any additional components in the breathing system Therefore: and any components that do not belong to the The condition of the patient and the ventilator's standard hose system can substantially increase measured values for air volume must be inspiratory and expiratory breathing resistance checked more frequently.
Preparation Connecting a hose system WARNING Make sure that the selected hose system cor- responds to the device setting! Carina might not give an alarm when using an incorrect hose system. In such cases, meas- ured volumes can differ significantly from the actual volumes! WARNING Do not use antistatic or conductive hoses.
Preparation Hose system with leak valve 1 Open the sliding cover on the underside of Carina. 2 Using a coin, turn the switch to leak valve » « until it is properly engaged. 3 Close the sliding cover. CAUTION Switching between a hose system with leak valve and a hose system with expiratory valve must be carried out by properly trained personnel only! Risk of malfunction!
Preparation 4 Connect the filter to socket » «. 5 Connect the ventilation hose to the filter. 6 Connect the expiratory valve to the nozzle of the ventilation hose. When using the disposable hose system VENTSTAR Carina ExpV (MP 00 313) the expiratory valve is included.
Preparation Connecting a respiratory gas humidifier WARNING Install a water trap when using a respiratory gas humidifier. Otherwise the flow inside of the hoses can be blocked by condensed water. WARNING Place the device and the respiratory gas humidifier on a stable, firm surface. Otherwise the humidifier might fall.
Preparation Respiratory gas humidifier without inspiratory heater For hose system with leak valve Turn the switch to leak valve » «, see page 41. 1 Connect the filter to socket » «. 2 Connect ventilation hoses, observing hose 60 cm lengths! 60 cm 3 Place the water trap vertically at the lowest...
Preparation Respiratory gas humidifier with inspiratory heater For hose system with leak valve Turn the switch to leak valve » «, see page 41. 1 Connect the filter to socket » «. 2 Fit the ventilation hoses according to the instructions for the respiratory gas humidifier, observing hose lengths.
Preparation Preparing the Aeroneb Pro nebuliser WARNING During medication nebulisation do not use a heat and moisture exchanger (HME). Risk of increased breathing resistance. WARNING Do not nebulise flammable medication, fire hazard! Position the nebuliser near to the patient. Observe the Instructions for Use of the neb- uliser being used! Instructions for Use Carina SW 3.n...
Preparation Connecting to the power supply CAUTION Maintain a sufficiently charged external battery on hand, in the possible event of a power supply failure. NOTE Route hoses safely to reduce the risk of tripping. A stationery device can be supplied either from the mains or from an external battery.
Preparation External battery connection 1 Insert the DC connector in the Carina. 2 Connect the external battery. Refer to the Instructions for Use of the external battery. External battery The external battery is not charged from the Carina mains supply! External battery for 3rd party connection 12 to 24 V DC (–10 % to +30 %);...
Preparation Internal battery If the external power supply (mains/external bat- tery) should fail, Carina automatically switches over to supply from the internal battery. If the battery is fully charged, ventilation can be per- formed reliably for approx. 60 minutes without any external power supply (assuming average ventila- tion parameters and normal ambient conditions).
Preparation supply, connect WARNING WARNING Keep oil and grease away from all oxygen sup- Make sure there is adequate ventilation at the ply components. Risk of explosion due to rear of Carina. Do not use an O source which spontaneous combustion! supplies a flow in excess of 10 L/min.
Preparation HPO – High Pressure Oxygen mode For O supply to Carina from the central gas supply or from a O gas cylinder. WARNING compressed gas must be dust-free, oil-free and dry and its pressure should be between 2.7 and 6 bar (270 and 600 kPa). Risk of fire. WARNING Only use O gas supplies with 100 % O...
Preparation Connecting to the nurse call system Socket on the back of Carina to transmit alarms to a central alarm system within the clinic. WARNING A nurse call system used in conjunction with Carina must conform to IEC 60601-1-8 and transmit alarms from Carina within 5 seconds.
Preparation MEDIBUS interface CAUTION The data port (data transfer) may only be con- nected by EDP (Electronic Data Processing) staff trained in maintenance and repair. Use Dräger "Babylink Cable" 83 06 488. Insert the plug into the » « socket on the rear of Carina and screw tight.
Page 54
This page intentionally left blank Instructions for Use Carina SW 3.n...
Start-up Information on start-up Before starting up for the first time and after any lengthy period of storage always charge the inter- nal battery of Carina for at least 2 hours. CAUTION The Carina shall not be operated immediately fol- lowing storage or transport outside the recom- mended operating conditions.
Start-up The self test is followed by the screen showing the ventilation mode and parameters last set. The device is in standby mode. Checking readiness for operation This test must be conducted every time before the device is put into operation. The checklist on page 58 can be copied for docu- mentation purposes.
Start-up Checklist Action Observe Check Connect all necessary equipment: hoses, Equipment is connected correctly, refer to cable, filter, etc. page 31. Turn the main switch on. Carina performs a self-test, LEDs light up and alarm tone is given. Check if the used hose system (leak valve or »LeakV«...
Start-up Switching Carina to Standby Press »Start/Standby« button and confirm with rotary knob. Carina is in standby mode. LED in the button flashes. The screen displays »Standby – Patient not ventilated«. Selecting an application method In the configuration menu set the method to »Mask«...
Start-up Setting the O supply Carina has been set to HPO mode by the manufac- turer. For LPO mode, configure LPO, see page 78. Dräger recommends the use of an external oxygen monitor add to the patient circuit to measure the actual oxygen concentration.
Operation Operation Setting ventilation modes ....62 Chart for operation with leak valve (MP 00 312)......79 Calling up the ventilation settings menu .
Operation Setting ventilation modes Switching on, Carina shows the ventilation mode last set along with the set parameters. Calling up the ventilation settings menu 1 Press »Select Menu« button. 2 Press the »Vent.Set« button and the ventilation mode last set is indicated. Changing the ventilation mode 3 Press button for the ventilation mode.
Operation Setting ventilation parameters 1 Press button for the ventilation parameter required. Select the value using the rotary knob and con- firm. Values set on delivery Ventilation mode PC-AC 500 mL 15 mbar insp PEEP 3 mbar 40 % 15 bpm 1.3 s Ti max Δ...
Operation VC-SIMV AF Volume Control – Synchronized Intermittent Man- datory Ventilation AutoFlow Volume-controlled ventilation with fixed mandatory Ramp high Ramp low minute volume MV, set with tidal volume VT and frequency f. Between the mandatory ventilation PEEP strokes the patient can breathe spontaneously. AutoFlow –...
Operation Setting ranges Ventilation parameter Setting range Tidal volume VT [mL] 100 to 2000 PEEP [mbar] 3 to 20 Inspiratory O concentration FiO 21 to 100 Trigger sensitivity Trigger Off/Normal/Sensitive Frequency f [bpm] 5 to 50 Inspiration time Ti [sec] 0.3 to 8 or I:E 1:39 to 2:1...
Operation PC-BIPAP Pressure Control – Biphasic Positive Airway Pressure Pressure-controlled ventilation in conjunction with Ramp high Ramp low free spontaneous breathing during the entire Pinsp breathing cycle. The mandatory portion of total PEEP minute volume MV is adjusted with inspiration pres- sure P , PEEP and frequency f.
Operation Setting ranges Ventilation parameter Setting range Inspiration pressure P [mbar] LeakV: 5 to 40 insp ExpV: 5 to 50 PEEP [mbar] 3 to 20 Inspiratory O concentration FiO 21 to 100 Trigger sensitivity Trigger Off/Normal/Sensitive Frequency f [bpm] 5 to 50 Inspiration time Ti [sec] 0.3 to 8 or I:E...
Operation PC-AC Pressure Control – Assist Control Pressure-controlled, assisted ventilation Ramp high Ramp low Every inspiratory effort of the patient on PEEP level Pinsp triggers a synchronized mandatory breath. PEEP Trigger window Flow with spontaneous without spontane- breathing ous breathing Set PC-AC with the buttons for the ventilation parameters: –...
Operation Setting ranges Ventilation parameter Setting range Inspiration pressure P [mbar] LeakV: 5 to 40 insp ExpV: 5 to 50 PEEP [mbar] 3 to 20 Inspiratory O concentration FiO 21 to 100 Trigger sensitivity Trigger Off/Normal/Sensitive Frequency f [bpm] 5 to 50 Inspiration time Ti [sec] 0.3 to 8 or I:E...
Operation SPN-PS Spontaneous Pressure Support For patients with sufficient spontaneous breathing. Spontaneous breathing at an increased pressure level to raise the functional residual capacity. Spon- taneous breathing can be supported with Pressure Support. Set SPN-PS with the buttons for the ventilation parameters: –...
Operation Setting ranges Ventilation parameter Setting range Pressure support Δ P [mbar] LeakV: 2 to 40 (with PEEP) supp ExpV: 2 to 50 (with PEEP) PEEP [mbar] 3 to 20 Inspiratory O concentration FiO 21 to 100 Trigger sensitivity Trigger Normal/Sensitive Maximum inspiration time Ti max [sec] 0.4 to 10.0...
Operation SPN-CPAP Spontaneous Continuous Positive Airway Pressure Suitable for patients with sufficient spontaneous breathing. Spontaneous breathing at an increased pressure level to raise the functional residual capacity. Set CPAP with the buttons for the ventilation parameters: – Positive end expiratory pressure »CPAP« –...
Operation Apnoea ventilation For automatically switching to volume-controlled, Start of apnoea Ventilation Spontaneous mandatory ventilation (VC-SIMV) in the event of breathing with apnoea. pressure support Can be switched on in ventilation modes SPN-PS PEEP and SPN-CPAP. The device gives an alarm »Apnoea ventila- tion!!!«...
Operation Volume guarantee Can be switched on in ventilation mode SPN-PS. Tidal volume guarantee Inspiratory plateau pressure is automatically regu- lated between P and PEEP in such a way that the set tidal volume VT is applied. Set volume guarantee with the parameters: –...
Operation NIV – Non-invasive Ventilation WARNING An intubated patient must not be ventilated in Mask ventilation is only possible in combination »Mask« mode! with a hose system with leak valve. A leakage may lead to a temporary change of Using mask ventilation the delivered volume to the patient.
Operation In the »Mask« application method all the ventilation modes are available for selection. Selecting the »Mask« application method Switch Carina to standby mode. Press »Select Menu« button. Press »Config.« button. Press » « button. Press »Method« button. Select and confirm »Mask« with rotary knob. The status line shows »Mask«.
Operation LPO – Low Pressure Oxygen mode NOTE WARNING If O is simultaneously supplied from the central Only connect O sources which meet the fol- gas supply or a gas cylinder and from the O con- lowing conditions: centrator, LPO supply is prevented. flow: 10 L/min max.
Operation Setting LPO mode Switch Carina to standby mode. Press »Select Menu« button. Press »Config.« button. Press » « button. Press »LPO« button. Select and confirm »On« with rotary knob. The status line on the screen shows »LPO«. Setting O concentration Set oxygen flow at the flow control valve in accordance with medical prescription:...
Operation Chart for operation with expiratory valve (MP 00 313) as a function of minute volume and LPO flow Minute volume in L/min Chart for operation with leak valve (MP 00 312) as a function of P and LPO flow mean Mean airway pressure in mbar Instructions for Use Carina SW 3.n...
Operation Displaying set values and measured values Measured values, set values and pressure curve are displayed. 1 Press »Select Menu« button. 2 Press »Values« button (only available during operation). The screen displays: – Settings – Measured values – graphic representation of Paw NOTE Measured Frequency on VC-SIMV and PC-BIPAP displays only mandatory breaths per...
Operation Terminating ventilation Press the »Start/Standby« button and confirm with rotary knob. Carina has terminated ventila- tion and is in standby mode. LED in the button flashes. The screen shows »Standby – Patient not ventilated«. Switching off the device Switch Carina to standby. 1 Set the mains switch to »...
Operation Improve displayed accuracy of the charging capacity of the internal battery NOTE If charging and discharging cycles are incom- plete, the charging capacity of the internal battery may be displayed inaccurately. If the internal indication accuracy is not reliable enough the following message will appear: »Battery indication not reliable !«...
Monitoring Setting alarm limits 1 Press »Select Menu« button. 2 Press »Alarms« button. Carina displays the alarm limits set. = Upper alarm limit = Lower alarm limit The current measured value is displayed between the upper and lower alarm limits. To set: Press the appropriate button.
Monitoring Extreme settings Certain ventilation parameters are limited by Carina to a threshold value. Extreme values can only be set after confirmation by pressing the rotary knob. In the event of an alarm If the alarm message, e.g. »MV high !!«, appears in the alarm line on the screen.
Monitoring More than one alarm at the same time If more than one alarm occurs at the same time, the alarm with the higher priority is displayed. When the cause has been remedied, the alarm with the lower priority is displayed. Troubleshooting Use the "Fault –...
Configuration Carina Configuration The following settings for application can be made under »Config.«: – Volume and type of alarm tone – Language, date and time – Screen display of curves and measured values – LPO mode on/off – Selection of mask or tube ventilation –...
Configuration The configuration menu shows the configurable functions and parameters: – Service for system and contact information and log book – Volume – adjustable from 1 to 5 – Tone – default tone or Dräger tone can be selected – Screen – adjustable for displaying curves or val- –...
Configuration Carina displays the functions of the Service menu: – System Information – for displaying device data and for setting system time and the current lan- guage – Contact – for displaying contact information for the customer – Log book – for displaying technical alarm mes- sages Displaying system information Press »Syst.info«...
Configuration Setting the language The language can be changed only on standby set- ting. On the »system information« screen, select the »current language« line by rotating the knob and confirming. The language is high- lighted. Set the appropriate language using the rotary knob and confirm.
Configuration Select Dräger tone or the default tone and con- firm. The tone selected is displayed. Selecting screen display The screen display can be set: 1 Press »Screen« button. Select »Curves« or »Measure« and confirm. The current setting is displayed in the button. 2 Pressure curve as a bar chart, 4 measured val- ues and 2 configurable measured values.
Configuration Selecting display for 2 measured values Value 1 and value 2 are freely selectable for the dis- play. 1 Value 1: Off, PIP, VTi (VT), f, PEEP, MVi (MV), leak mean 2 Value 2: Off, PIP, VTi (VT), f, PEEP, MVi (MV), leak mean The measurements for value 1 and value 2 cannot...
Configuration Selecting Ti or I:E Set I:E (time ratio of inspiration to expiration) or inspiration time Ti. 1 Press »Ti / I:E« button. Select »Ti« or »I:E« and confirm. Setting the night screen If the device is set to »Night« screen and it is not operated for 3 minutes or no alarm has occurred, the screen switches off.
Fault – Cause – Remedy Fault – Cause – Remedy Carina displays alarm messages in hierarchical order in the alarm line on the screen. If two faults are detected at the same time, the more critical fault is displayed. The priority of alarm messages is indicated by dif- ferent background colours and with exclamation marks after the message: !!! =...
Page 97
Fault – Cause – Remedy Error message Cause Remedy Battery indication not reliable Battery indication is not relia- Proceed the action on page 82. ble. If the problem persists call DrägerService. Blower overheated The blower has overheated. Make sure air is being supplied (in log book »Blower over- The device is operating at to the device.
Page 98
Fault – Cause – Remedy Error message Cause Remedy External power failure The device has no external Check the mains supply or DC power supply. power supply. The mains plug or DC plug is Insert the mains plug or DC loose.
Page 99
Fault – Cause – Remedy Error message Cause Remedy Int. battery low Battery charge has dropped to Charge the battery: connect below 25 %. external power supply. Int. battery low Battery charge has dropped to Charge the battery: connect below 50 %. external power supply.
Page 100
Fault – Cause – Remedy Error message Cause Remedy MV high The minute volume has Check condition of patient, exceeded the upper alarm check ventilation pattern, cor- limit. rect alarm limit if necessary. Leakage in breathing system. Ensure that breathing system is leak-proof.
Fault – Cause – Remedy Error message Cause Remedy Standby activated The device is ready for opera- To start ventilation: press the tion. Ventilation is in standby »Start/Standby« button. mode. Technical alarm The device has a technical Use a replacement device. (in log book problem.
Page 102
This page intentionally left blank Instructions for Use Carina SW 3.n...
Cleaning, Disinfection and Sterilisation Cleaning, Disinfection and Sterilisation General information on cleaning, disinfection and sterilisation... . . 104 Care list ......105 Dismantling the components .
Cleaning, Disinfection and Sterilisation General information on cleaning, disinfection and sterilisation WARNING In case of potential contamination with body fluids, clean and disinfect the device. Wear protective clothing, goggles, disposable gloves etc. Risk of infection for hospital staff and other patients. WARNING Personnel should take precautions to avoid direct contact with chemical disinfectants and...
Cleaning, Disinfection and Sterilisation Care list What How often Cleanable components Recommended cleaning intervals Carina basic device per patient Dismantle, clean and if applicable wipe disinfect Trolley per patient Dismantle, clean and if applicable wipe Compressed gas hose disinfect Hinged arm Reusable ventilation hoses per patient/weekly Dismantle, clean and disinfect in the...
Cleaning, Disinfection and Sterilisation Cleaning Wipe disinfection with Virex Tb Wipe surfaces with a clean, damp lint-free cloth. This procedure is applicable for Virex Tb. When Do not use corrosive detergents. using Virex Tb or other disinfectants strictly follow the disinfectant manufacturer's instructions. Aged or accumulated deposits should not be Clean the surfaces according to the cleaning scraped off, but should instead be softened with...
Maintenance Maintenance intervals Clean and disinfect equipment and/or compo- nents before any maintenance procedures and before returning for repair. Inlet filter (HEPA filter) Replace after 6 months or in accordance with hospital-specific regulations, page 109. For disposal see page 112. NiMH battery To be replaced by technicians after 3 years.
Maintenance Replacing the inlet filter WARNING Do not operate the device without an inlet fil- ter: risk of infection! When the inlet filter is removed the »Inlet filter missing !!« alarm is displayed. 1 Unlock the inlet filter, fold it down and remove it from the holder.
Maintenance Maintenance of the internal battery The battery has no memory effect. If the battery is not used regularly it discharges. Charge at regular intervals depending on ambient temperature: Every 3 months for –20 to +35 °C Every month for +35 to +55 °C Charge the battery more frequently at higher tem- peratures.
Disposal Disposal Disposing of the internal battery ..112 Disposing of the inlet filter ....112 Disposing of the device ....112 Instructions for Use Carina SW 3.n...
Disposal Disposing of the internal battery Do not throw onto a fire, risk of explosion! Do not open by force, risk of acid burns! NiMH batteries are special waste: dispose according to relevant national waste disposal regulations. For information consult your environment agency, local government offices or appropriate waste dis- posal companies.
Technical Data Ambient conditions During operation Temperature +5 to +35 °C Atmospheric pressure 900 to 1100 hPa reduced functionality at 700 to 900 hPa Rel. humidity 5 to 95 % non-condensing In storage and transportation Temperature (up to 1 day) –20 to +60 °C Temperature (up to 1 month) –20 to +55 °C...
Technical Data Inspiration time Ti 0.3 to 8 seconds Resolution 0.1 second Inspiration/expiration ratio I:E 1:39 to 2:1 Ramp rise time Ramp Auto (device uses the measured breathing rate to calculate the optimum value) 0.1 to 2.0 seconds Resolution 0.1 second Trigger sensitivity Trigger Off, Normal, Sensitive Apnoea ventilation Apn.
Technical Data Inspiratory tidal volume VTi, Range 0 to 4000 mL, BTPS Resolution 1 mL Accuracy ±10 % of the measured value (without leakages) Waveform display Airway pressure Pressure (t) 0 to 80 mbar Flow (t) –160 to +160 L/min Bar graph display Airway pressure Pressure 0 to 60 mbar...
Technical Data Operating data Power supply Mains 100 to 240 V AC (–20 to +10 %); 50 to 60 Hz; 1.7 to 1.1 A Internal battery Charging capacity 2.5 Ah Type NiMH Charging time 130 minutes Operating time 60 minutes (at average ventilation parameters and normal ambient conditions) Remaining operating time 3 minutes (time between loss of all power and generating of...
Technical Data Screen TFT color screen 13.7 cm (5.4 inch) Classification Classification in accordance with Direc- tive 93/42/EEC Appendix IX Protection class II, acc. to IEC 60601-1 Connection block Type BF Electromagnetic compatibility EMC tested according to IEC 60601-1-2: 2001 UMDNS code 14-355 Universal Medical Device Nomenclature...
Technical Data Hose systems Ambient conditions Temperature during operation 10 to 40 °C Temperature during storage –20 to 60 °C Atmospheric pressure 500 to 1100 hPa Rel. humidity 5 to 95 % non-condensing Performance Data Resistance at 60 L/min <2 mbar Compliance <1.5 mL/mbar Leakage...
Technical Data EMC Declaration General information The EMC conformity of Carina includes the use of the following external cables and accessories: Description Order-No. DC-cable S, 1.75 m 84 14 048 Cable mains Euro CL II, 2.5 m 57 30 042 Power cable Switzerland, 18 51 691 3 m, 10 A...
Technical Data Electromagnetic emissions Carina is intended for use in the electromagnetic environment specified below. The customer or the user of the Carina should assure that it is used in such an environment. Emissions Compliance according to Electromagnetic environment Radio frequency emission Group 1 Carina uses RF energy only for its (CISPR 11)
Page 122
Technical Data Immunity against IEC 60601-1-2 Compliance Electromagnetic environment test level level (of Carina) Power frequency mag- 3 A/m 3 A/m Power frequency magnetic fields netic field (50/60 Hz) should be at levels characteristic of a (IEC 61000-4-8) typical location in a typical commercial or hospital environment.
Technical Data Recommended separation distances Recommended separation distances between port- able and mobile RF-telecommunication devices and the Carina. Max. P 3 V/m distance 1 V/m distance (m) Examples EIRP 0.001 0.06 0.17 0.003 0.10 0.30 0.010 0.18 0.55 0.030 0.32 0.95 e.g.
Page 124
This page intentionally left blank Instructions for Use Carina SW 3.n...
Description The measured values of the first pressure sensor proportional valve (V5). Adding to the respiratory (PS1) and the second pressure sensor (PS2) are air takes place in the blower (2). used to regulate ventilation pressure. The calcula- tion incorporates the following: set values, ventila- Adding oxygen LPO tion algorithms and characteristic of the original The switchover to low-pressure oxygen is per-...
Description Compensation of volume and pressure losses Carina with hose system with leak valve compen- sates leakages from valve and patient. However, to compensate the volume loss due to the compliance of the different hose systems, con- sider the following table: ΔP above Hose system Volume loss...
Description VC-SIMV Volume Control – Synchronised Intermittent Man- Unsynchronised datory Ventilation Synchronised manda- mandatory tory ventilation stroke ventilation stroke Hybrid form of mandatory ventilation and spontane- ous breathing. Ramp Ramp SIMV enables the patient to breathe spontaneously in regular prescribed cycles, with the mechanical PEEP mandatory ventilation strokes providing a minimum Trigger-...
Description In the course of progressively weaning the patient from artificial ventilation, the ventilation frequency f is further reduced while the spontaneous breathing time is increased, so that the required total minute volume is supplied more and more by spontaneous breathing.
Description If the patient breathes in or out during mandatory inspiration, inspiration pressure PIP does not change for that ventilation stroke – only inspiration flow and expiration flow adapt to patient demand. The individually administered tidal volume VT may differ from set tidal volume VT in single ventilation PEEP strokes but the tidal volume VT administered on average over time, is constant.
Description PC-BIPAP Pressure Control – Biphasic Positive Airway Pressure This PC-BIPAP ventilation mode is pressure-con- Ramp high Ramp low trolled or time-controlled ventilation during which Pinsp the patient can always breathe spontaneously. PEEP PC-BIPAP is thus often described as a time-control- led change between two pressure levels.
Description The display of calculated tidal volume VT is used to set the necessary difference between the two pres- sure levels. As in VC-SIMV, the time pattern is set with the parameters of the basic setting for frequency f and inspiration time Ti.
Description SPN-PS Spontaneous Pressure Support For pressure support of insufficient spontaneous breathing. The function of the device in supporting insufficient Δ P spontaneous breathing is similar to that of the supp above anaesthetist who manually assists and monitors PEEP the patient's spontaneous breathing by feeling the Ramp breathing bag.
Description SPN-CPAP Spontaneous Continuous Positive Airway Pressure Spontaneous breathing at a raised pressure level in CPAP order to increase the functional residual capacity (FRC). Flow Tube (Invasive ventilation) The Carina is used to provide invasive ventilation to the appropriate patients. The following guidelines should be considered prior to use.
Description Using invasive ventilation with VentStar Carina ExpV (MP 00 313) This hose system has an expiratory valve. There- fore the exhaled gas runs through the expiratory valve instead of the inspiratory limb when the device is not operating. Mask ventilation (Non-invasive ventilation) NIV (the mask ventilation) WARNING...
Description Information concerning CO re-breathing WARNING In general, as inspiratory time increases, the WARNING potential for CO re-breathing increases. A Never leave the mask on patient while Carina higher inspiratory time decreases exhalation is not ventilating. There is a risk of CO time, allowing less CO to be purged from the breathing in case of no ventilation.
Description SyncPlus Volume Guarantee (VG) SyncPlus is an automatic function for supporting The ventilation mode extension VG allows for the NIV (mask ventilation). volume support on SPN-PS mode. Carina controls This automatic technology are supported by 3 main inspiratory plateau pressure automatically, in order features: to deliver the set tidal volume.
Description Volume measurement The gas volume that did not flow back through the sensor must have escaped through the leak. VTi (hose system with leak valve) Volume accuracy in case of leakages VTi value displayed on the screen is the volume of Due to Carina's limited monitoring functionality (no air delivered to the patient’s lungs (not in internal monitoring of expiratory minute volume),...
Description AutoRamp Ramp setting has 2 options – Time setting: The steepness of the pressure rise from the PEEP level to the inspiration pres- sure level can be even more closely adapted to the needs of the patient in all ventilation mode except CPAP.
Index Index connecting ......40 with expiratory valve ....41 Abbreviations .
Page 142
Index Rear ........19 Respiratory gas humidifier ....43 with inspiratory heater .
Page 143
This page intentionally left blank Instructions for Use Carina SW 3.n...
Page 144
These Instructions for Use only apply to Carina SW 3.n with the Serial No.: If no Serial No. has been filled in by Dräger, these Instructions for Use are provided for general information only and are not intended for use with any specific machine or device.