A 803 Holder For Transmission Instruments; Installation - Miele A 105/1 Operating Instructions Manual

Upper injector basket
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 33

Installation

Where necessary, the nozzles supplied can be replaced with different
fittings, e.g. holders for dental transmission instruments, irrigation
sleeves, etc. Further examples are given on the following pages.
 The wash water used for reprocessing narrow lumen instru-
ments needs to be filtered. The injector manifolds can each be fit-
ted with an A 800 tubular filter for this purpose. The tubular filter is
supplied with its own operating instructions.
Cleaning the A 800 tubular filter
The tubular filter must be cleaned at least once a week.
We recommend cleaning the tubular filter after every 10 programme
sequences.
Follow the cleaning instructions in the operating instructions for the
tubular filter.
A 803 holder for transmission instruments
A 803 holders must be retro-fitted for reprocessing dental transmis-
sion instruments. The holders are suitable for almost all standard
transmission instruments.
Remove a nozzle from the injector manifold and replace it with an
A 803 holder.
When the A 803 holders are used, the wash water must be filtered.
The injector manifold must be fitted with an A 800 tubular filter.
The tubular filter is supplied with its own operating instructions.
Follow in particular the instructions on how to perform regular
cleaning on the tubular filter.
en - Areas of application
43

Advertisement

Table of Contents
loading

Table of Contents