Illustration
NOTE: If readying the bed to admit a new patient, power down the warmer for 10 seconds,
then turn power back on. Allow warmer to preheat the bed in Warmup Mode.
2-6
Panda iRes Warmer
Step
6
Unplug the skin probe and change the warm-
er to Baby Mode. Check that both visual
and audio alarms trigger in the Baby Mode.
Silence the alarm by using either the alarm
silence key or the Hands Free Alarm silence.
7
Return the warmer to Manual Mode by press-
ing the "Mode" key. Select a Heater Power
setting to silence the prompt tone.
8
If the unit is equipped with an elevating bed,
check the operation of the bed elevation
mechanism. Raise and lower the bed along
its entire travel range, checking that the
mechanism operates smoothly. Check that
the two bed height adjustment pedals on
either side of the unit raise and lower the bed
height.