Maintenance

Replacing the Writer Roller/Carriage Assembly

Use the following procedure to replace the writer roller/carriage assembly.
1.
Remove the power supply following the procedure described in
Power Supply" on page
2.
Inside the power supply compartment, disconnect the cable connected to the
writer assembly.
3.
Open the unit to access the paper compartment and move the paper size
bracket to the A4 position to expose one of the writer assembly mounting
screws.
Refer to
4.
Remove the screw and return the paper size bracket to the 8.5 x 11 position.
5.
Close the unit and turn it over so the bottom side is up.
6.
Remove the four screws located on the underside of the writer roller/carriage
assembly and lift the writer from the bottom of the unit.
7.
Reverse step

Replacing the Barcode Reader Cable

Use the following procedure to replace the cable on a barcode reader.
1.
Position the barcode reader so the label (1) is in front.
2.
Carefully peel back the label (2) so the pin hole is revealed.