Replacing the Writer Roller/Carriage Assembly

Use the following procedure to replace the writer roller/carriage assembly.
1.
Remove the power supply following the procedure described in
Power Supply" on page
2.
Inside the power supply compartment, disconnect the cable that connects
to the writer assembly.
3.
Open the unit to access the paper compartment and move the paper size
bracket to the A4 position to expose one of the writer assembly mounting
screws.
Refer to
4.
Remove the screw and return the paper size bracket to the 8.5 x 11 position.
5.
Close the unit and turn it over so the bottom side is up.
6.
Remove the four screws located on the underside of the writer roller/carriage
assembly and lift the writer from the bottom of the unit.
7.
Reverse the previous steps to insert the new writer roller/carriage assembly.

Replacing the Trolley Casters

Use the following procedure to replace the casters on the MAC Series trolley.
1.
Remove the MAC unit and all loose items from the trolley and then place the
trolley on its side.
2.
Locate the slot under the arrow on the bearing dust cap and, using a small blade
screwdriver, pry the cap from the caster to be removed.
2046275-017G
103.
"Setting the Correct Paper Size" on page 95
MAC™ 5500/MAC™ 5500 HD
Maintenance
"Replacing the
for details.
127