6 LCD DISPLAY PANEL

The LCD display panel is located on the inner cover and is visible when the door is open
as shown in Figure 6—1.
Figure 6—1: Inside view
During an analysis, the display panel shows the serial number of the product and the
operational state. The last measurement results with date and time stamps can be
obtained by using the black scroll buttons to page up or down through the details (see
Section 6.2 for examples). By default, automatic paging is turned off.
In the event of an error, details are shown on the display panel (see Figure 6—2 for an
example). The middle line indicates the state within the measurement process at which
the error occurred — denoted by a double-digit code. For example, 02 indicates an error
in the oil purging state. The last line indicates the errors in that state — denoted by a
sequence of double digits codes each representing a specific error. For example, 18
means that the oil pressure is too low (see Section 6.2 for a list of defined error codes).
Note: If contacting the Customer Service Centre, ensure that you provide the state code and error code.
DGA500
>>>> ERRORNN <<<<
NN, NN, NN, NN, NN, NN, NN
DGA500
>>>> ERROR02 <<<<
18
Figure 6—2: LCD information notice – error format (top) & example (below)
MA-014 - TRANSFIX DGA 500 User Guide – Rev 1.1
Display Panel
Scroll Up
Scroll Down
Start
Stop
USB Connection
State code
Error codes
13-Oct-15
Page 12 of 28