A B B M E A S U R E M E N T & A N A LY T I C S | O P E R AT I N G I N S T R U C T I O N | O I/ 2 6 6 H S H M T- E N
266HSH Modbus Transmitter
Pressure/Temperature Multivariable
Introduction
266 models
The 2600T family provides comprehensive range of
top quality pressure measurement products,
specifically designed to meet the widest range of
applications ranging from arduous conditions in
offshore oil and gas to the laboratory environment
of the pharmaceutical industry.
Engineered solutions for all
applications
Measurement made easy
For more information
Further publications for 2600T series pressure
products are available for free download from
www.abb.com/pressure
Data Sheet
Product
Operating Instruction
Product
Search for or click on:
DS/2101130-EN
OI/266HSHMT-EN

Advertisement

Table of Contents
loading

Summary of Contents for ABB 266HSH

  • Page 1 A B B M E A S U R E M E N T & A N A LY T I C S | O P E R AT I N G I N S T R U C T I O N | O I/ 2 6 6 H S H M T- E N 266HSH Modbus Transmitter...
  • Page 2: Table Of Contents

    Table of contents 1 Introduction ............3 6.10 Correcting the zero shift ........... 15 6.11 Installing/Removing the LCD display ...... 16 1.1 Instruction manual structure ......... 3 6.12 Securing the housing cover in flameproof areas ..16 1.2 Models covered by this manual ........3 1.3 Product description ............
  • Page 3: Introduction

    1.2 Models covered by this manual The present manual can be used for the Modbus 266 For safety reasons, ABB draws your attention to the fact that only pressure model. sufficiently insulated tools conforming to EN 60900 may be used.
  • Page 4: Warranty Provision

    Proper disposal prevents negative effects on people and the environment, and supports the reuse of valuable raw materials. ABB can accept and dispose of returns for a fee. Attention – <Property damage>. This message indicates a potentially damaging situation. Failure to avoid this could 2.11 Transport and storage...
  • Page 5: Transmitter Overview

    Corrective maintenance work may only be performed by trained personnel. — Before removing the device, depressurize it and any adjacent lines or containers. — Check whether hazardous materials have been used as materials to be measured before opening the device. Residual amounts of hazardous substances may still be present in the device and could escape when the device is opened.
  • Page 6: Range & Span Consideration

    3.2 Range & Span consideration The 2600T Transmitter Specification Sheets provide all information concerning the Range and Span limits in relation to the model and the sensor code. The terminology currently used to define the various parameters is as follows: URL: Upper Range Limit of a specific sensor.
  • Page 7: Optional Wired-On Sst Plate (I1)

    Calib. Range 2600T Number PRESSURE TRANSMITTER Figure 2: Product identification C and F sensors on gauge and absolute pressure transmitters are in compliance with IEC61000-4-6 with B criteria 4.2 Optional wired-on SST plate (I1) 4.4 Storage The 266 transmitter can be supplied with the optional The instrument does not require any special treatment if “Wired On Stainless Steel plate”...
  • Page 8: Mounting

    Before installing the transmitter, check whether the device required in normal condition. ABB typically configures 266 design meets the requirements of the measuring point pressure transmitters according to the user requirements.
  • Page 9: B6 & B7 Barrel Housing Bracket

    2 – U-bolt fixing washers and nuts 3 – Transmitter fixing bolts 4 – B6 or B7 bracket 5 – Fitting adapter (supplied with 266HSH) Figure 5: Pipe and wall mounting bracket kits for P style transmitter with Barrel housing 128 (5.04) 29 (1.14)
  • Page 10: Din Housing Bracket Details

    128 (5.04) 29 (1.14) 120 (4.72) 30 (1.19) 18 (0.71) 83 (3.27) 18 (0.71) 66 (2.60) 72 (2.83) 1/2 - 14 NPT 105 (4.13) 22 (0.87) width across ats of exagon 117 (4.60) Figure 7: Model 266G or 266A P-Style transmitter with DIN housing installed on a 2”pipe with optional bracket (B7 Stainless Steel 316L) 5.4.4 B7 DIN Housing bracket details 1 –...
  • Page 11: Setup

    6 Setup NOTICE If cable glands are not used, the red transport screw plugs must be replaced with suitable screw plugs when 6.1 Temperature measurement the transmitter is installed. This is because the transport — Mount the temperature sensor in the downstream pipe screw plugs are not certified as protected against of the primary element.
  • Page 12: Grounding

    65 x 10 + 10000 Terminal Function/comment – Power supply R x C PWR+ Where: PWR- A (+) Modbus interface RS485 L = Line length in meters B (-) R =  Total resistance in Ω (ohms) 1, 2, 3, 4 Terminals for connecting the Pt100 resistance thermometer C = Line capacitance in pF/m NOTICE...
  • Page 13 Cable specification Using the Modbus protocol allows devices made by different manufacturers to exchange information via the The maximum permissible length depends on the baud same communication bus, without the need for any rate, the cable (diameter, capacity and surge impedance), special interface devices to be used.
  • Page 14 Figure 13: To connect the pressure transmitter, observe the NOTICE following instructions: If the O-ring gasket is seated incorrectly or damaged, this — Lead the cable for the power supply and the may have an adverse effect on the housing protection class. Modbus connection into the terminal box.
  • Page 15: Write Protection

    (refer to chapter 9.6), external, non-intrusive push buttons can be installed (if not selected with the option R1) by ordering the commercial code DR1014. Please refer to local ABB representative. Set Lower Range Value — Apply the lower range value pressure from the process or from a pressure transducr.
  • Page 16: Installing/Removing The Lcd Display

    7 Operation 6.11 Installing/Removing the LCD display —Unscrew the housing cover of the communication 7.1 Local push buttons functionality (option R1) board/LCD side. 266 transmitters allow local adjustments via the on-board Important. With an Ex d / Flameproof design, please refer to non intrusive push buttons, when selected.
  • Page 17: Configuring The Transmitter Without An Integral Lcd Hmi

    7.4 Configuring the transmitter without an 7.5 LRV and URV configuration integral LCD HMI — Apply the pressure for the “lower range value” and wait approx. 30 s until it has stabilized. The “lower range value” and “span” parameters can be set directly on the transmitter using the external or internal —...
  • Page 18 Replace mode (DIP switches 1 and 2) NOTICE In normal mode the DIP switches 1 and 2 are in position 0. Always disconnect the device from the power supply If a replacement procedure is necessary, they will be before making changes to DIP switches. activated.
  • Page 19: Lcd (L1 And Ls Option) Activation Considerations

    Figure 19: Display keypad of L1 and L5 LCD display options The keys (1) , (4) , (2) and (3) are available for the menu-controlled configuration. — The menu/submenu name is shown at top of LCD display. — The number/line of the currently selected menu item is displayed in the upper right of the LCD display. —...
  • Page 20: Configuration Of The Pressure Transmitter Using

    7.10 7.8 Configuration of the pressure transmitter Important. After 30 seconds without action on on-board buttons or using the optional LCD HMI with keypad (menu- external Z/S pushbuttons, the Standard HART transmitter controlled) automatically exit the configuration menu. The integral LCD option (L1 or L5 option) is connected on the 266 Advanced HART communication board.
  • Page 21 Menu Display Exit Select This menu allows the set up of different functions rele- This menu gives you all information about the device. The vant to the display itself. The menu driven structure will menu driven structure will show you what is the sensor guide you through the choice of some functional aspects type, the hardware and software revisions, the high and low Display...
  • Page 22 Min -14.504 0.0000 Next Edit Next Easy Setup PV Upper Range Value 6000.0 psi Upper Range Value Max 8702.3 Easy Setup Menu Min -14.504 Easy Setup In the configuration level, use the (2) + (3) Next Edit Next keys to scroll to the “Easy Set-up” menu Easy Setup Linearization Type item and select it by pressing the (4) key .
  • Page 23 Simply press “ok” (4) key to activate the PV to zero function. Press Next Max 8702.3 Set PV to Zero Min -14.504 0.0000 (1) key to move to the next menu item. ABB suggest user should perform this specific command only after the installation and Next Edit Next...
  • Page 24 Menu Device Setup - Output Scaling Device Setup - SW Write Protection - Factory Reset - Process Variable - Transfer Function Exit Select Device Setup Device Setup Sw Write Protection Sw Write Protection Unlocked Sw Write Protection Process Variable Locked Transfer Function Unlocked Output Scaling...
  • Page 25 Process Variable Apply PV Apply PV PV Type Lower Range Value Lower Range Value Set PV Upper Range Value Apply PV Parallel Shift Back Select Back Select Back Apply PV Apply PV Lower Range Value Upper Range Value Upper Range Value Back Select Back...
  • Page 26 Device Setup Output Scaling Output Scaling Output Scaling Damping Sw Write Protection Damping Process Variable Unit Transfer Function Lower Range Value 0.0000 Min 0.0000 Upper Range Value Output Scaling Back Select Back Select Back Select Next Output Scaling Unit Unit Damping Pressure Pressure...
  • Page 27 Level Cancel Unit Unit Mass and Volume Other Density Level Other Back Select Back Edit Other Cancel Output Scaling Output Scaling Lower Range Value Damping Lower Range Value Unit Max 999999 Lower Range Value Min -999999 Upper Range Value Back Select Back Edit...
  • Page 28 Display Menu Menu - Line 1 View - Language Display - Line 2 View - Display Revision - Bar View - Contrast -Mode Exit Select Display Display Language Language English Language Display Revision Deutsch Contrast Italiano English Mode Back Select Back Edit Cancel...
  • Page 29 Display Display Scaling Display Linearization Type Line 1 View Linearization type Linear Linearization Line 2 View Unit Square Root Bar View Lower range value Sqrt. 3rd power Linear Display Scaling Upper range value Sqrt. 5th power Back Select Back Select Back Edit Back...
  • Page 30 Menu Calibrate Calibrate - Main Pressure Sensor - Temperature Sensor Exit Select Calibrate Unit Main Pressure Sensor Low Trimming Main Pressure Sensor 0.000 Main Pressure Sensor Main Pressure Sensor Main Pressure Sensor Low Trimming Low Trimming Temperature Sensor High Trimming Max 60000 -100.00 Next...
  • Page 31 Menu Diagnostics Diagnostics - Monitoring Exit Select Diagnostics Monitoring Monitoring Monitoring Pressure Pressure Output % Scaled Output Static Pressure Back Select Back Select Back Monitoring Monitoring Pressure Output % Output % Scaled Output Static Pressure Back Select Back Monitoring Monitoring Pressure Scaled Output Output %...
  • Page 32 Menu Device Info Device Info - High Sensor Limit - Hardware Revision - Low Sensor Limit - Software Revision - Minimum Span Exit Select Device Info Device Info High Sensor Limit High Sensor Limit Low Sensor Limit Minimum Span 60000kPa Hardware Revision Back Select...
  • Page 33: Damping (Damping)

    Menu Communication Communication - Protocol Settings - Protocol - HART Address Exit Select Communication Communication Protocol Settings Protocol Settings Protocol HART Address Software Back Select Back Communication Communication Communication Protocol Settings Operate/Modbus Protocol Protocol HART Address Back Select Back Back Select Communication HART Address...
  • Page 34: Configuration With Teh Pc/Laptop Or Handheld Terminal

    DHH800-MFC, Emerson Process 375 and 475 provided the 266 EDD has been downloaded and enabled in the terminal. — ABB Asset Vision Basic, a new free of charge software configurator downloadable at www.abb. com/Instrumentation. — Any DTM based software for HART instruments configuration provided it is compatible with EDD or DTM.
  • Page 35: Configuration With The Graphical User Interface (Dtm) - System Requirements

    — unit strings on display (Nm3, bbl, t) 01/2013 interface (DTM) - System requirements were wrong — Operating control program (e.g., ABB Asset Vision New Features: Basic version 1.00.17 or higher) — new language (French & Spanish) on local indicator —...
  • Page 36: Standard Hart Software Revision History

    7.16 Standard HART software revision history 7.17 Revision Description Release date From 7.1.50 First release 02/2013 7.1.51 Internal release not published 7.1.52 Internal release not published Bug Fixing: — HART burst mode — correcting Custom unit visualization on LCD 7.1.52 7.1.53 Improvements: 01/2014...
  • Page 37: Error States And Alarms

    8.2 Error states and alarms — Communication Board / Electronic related error messages. Error message Tx LCD message Possible cause Suggested action Tx response Electronic Memory F116.023 Electronic memory corrupted. The electronic must be replaced. Analog Signal to Alarm Failure A DAC (digital to outpt converter) trimming The output circuit could be Output ReadBack...
  • Page 38 — Configuration related error messages. Error Tx LCD message Possible cause Suggested action Tx response message Use a HART configurator (DTM - Hand held) to Input Simulation The P-dP Value produced in output is C088.030 place device back in to normal operating mode no effect Active derived by the value simulated in input.
  • Page 39 — Operation related error messages Error message Tx LCD message Possible cause Suggested action Tx response Check the Voltage at the terminal block and if it is Power Supply The Device Power Supply is close to the lowest M024.036 not within the valid range check the external no effect Warning acceptable limit.
  • Page 40: Maintenance

    The use of non original spare parts makes the warranty expected to accumulate, the measuring equipment should void. In case you want ABB to perform the repair, please be cleaned on a regular basis, in accordance with the send back the transmitter to your local ABB office operating conditions.
  • Page 41 — Respect the below table indications for reinstalling the process flanges. Transmitter model and range Procedure Viton Gaskets All bolting Use a torque wrench to tighten the bolts to a torque of 25 Nm. Carbon Steel NACE Use a torque wrench to tighten the process flange nuts to a torque of 40 Nm, let the flange 266DSH / PSH / VSH and Stainless Steel stabilize for an hour, unscrew the nuts and tighten again to 25 Nm.
  • Page 42: Hazardous Area Considerations

    seconds and lower dip-switch 1 to 0 position. 266 can Important. The number close to the CE marking of the reconfigure itself with the previous configured transmitter safety label identifies the Notified Body which has parameters thanks to the auto-configuration responsibility for the surveillance of the production.
  • Page 43 b) Certificate ATEX II 1/2 G Ex ia IIC T4/T5/T6 Ga and II Ex ia IIIC T85°C Da 1/2 D Ex ia IIIC T85°C Da IECEx certificate number FM Approvals certificate number IECEx FME 16.0003X (Tremezzina, Minden, Bangalore, FM09ATEX0024X (Tremezzina, Minden, Bangalore and Shangai products) Shanghai products) The other marking refers to the protection type used...
  • Page 44 c) Certificate ATEX II 1/2 G Ex db IIC T6 Ga/Gb and II 1/2 D Certificate IECEx Ex db IIC T6 Ga/Gb and Ex tb IIIC T85°C Ex tb IIIC T85°C Db Db, Ta= -50°C to +75°C - IP67 Ta = -50°C to +75ºC – IP67 IECEx certificate number FM Approvals Certificate number IECEx FME 16.0002X (Tremezzina, Minden and Shanghai...
  • Page 45 According to ATEX Directive (European Directive 2014/34/ Important. It is the technical support for the ABB Declaration EU and relative Standards which can assure compliance with of Conformity. Essential Safety Requirements, i.e., EN 60079-0 (General requirements) EN 60079-15 (Specification for electrical Important.
  • Page 46: Entities For "L5" Option

    10.1.1 Entities for “L5” option E, F and G, hazardous (classified) locations. (display with TTG technology) — Suitable for Class II, III, Division 2, Groups F and G, HART Version with “L5” option (display TTG) hazardous (classified) locations. Ui= 30Vdc Ci= 5nF Li= uH Temperature Temperature Minimum...
  • Page 47 We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.