Miele A 204 Operating Instructions Manual page 33

Mobile injector unit
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 25
Screw long hollow instruments to flexible hose connectors using 
Luer-Lock adapters.
When processing hollow instruments with several connectors, such
as rinsing and suction hand pieces, each connector needs to be se‐
cured to a hose.
Hollow instruments with a  connector should be connected to the
hose using an E 790 Luer lock  / Luer lock  connector.
Fit hollow instruments securely into the silicone holders on the rack
in the injector unit.
Space the silicone holders in the rack according to the size of the
instruments.
Unscrew hollow instruments after successful reprocessing.
 Any hollow items that have become disconnected from their in‐
jector nozzles, adapters or holders during processing must be re-
processed.
en - Areas of application
33