Setting VCR (Mitsubishi MD3000) USB Interface

1.) Turn off the Power of the VCR.
2.) Open the rear panel of the VCR.
3.) Set dip switches. Off 1~6 pins, On 7 pin, and Off 8 pin.
NOTE:
7 Pin ON = USB Interface, OFF = RS232C Interface.
Figure 3-15 Setting VCR USB Interface